Model MG-III - Power Factor Test Set
Measures the dissipation factor (Tan Delta, power factor), relative permittivity (dielectric constant), capacitance, insulation resistance and volume resistivity of electrical insulating oils and other insulating liquids. Conforms to IEC 60247, ASTM D924, IS 6262/6103, DL 474.3 and GB/T 5654. CE Mark.
-
Most popular related searches
Features
- Microprocessor controlled inductive heater - uniformity of temperature throughout the cell;
- Automatic measurement;
- High accuracy;
- Built-in thermal printer for test report printing;
- Large LCD touch screen.
Technical Specifications
- Oil cup heating up to 100°C with uniformity across the cup (For routine testing, usually at 25°C, 85°C or 100°C)
- Output voltage: 1000V - 2200V (200V interval)
- Tan Delta measurement range: 0.000% - 100.0%
- Accuracy: ±1.0% of reading + 0.00001
- Capacitance measurement range: 15pF - 300pF
- Accuracy: ±1.0% of reading + 0.01pF
- Resistance measurement range: 10Mohm - 10Tohm
- Accuracy: ±10% of reading
- Relative permittivity and Volume resistivity calculated automatically and displayed.
- Capacitance of empty test cell: 60pF (±2pF)
- Ambient temperature: -5°C - 40°C
- Power supply: AC 220V, 50Hz