Sigma-Aldrich
Gas Generators
Supelco Park
Bellefonte, PA 16823-0048 USA
Telephone 800-247-6628 • 814-359-3441
Fax 800-447-3044 • 814-359-3044
email: supelco@sial.com
http://www.sigma-aldrich.com
S u p e l c o i s a M e m b e r o f t h e S i g m a - A l d r i c h F a m i l y ,
P r o v i d i n g C h r o m a t o g r a p h y P r o d u c t s f o r A n a l y s i s a n d P u r i f i c a t i o n .
Bulletin 898C
T196898C
AYW ©2000 Sigma-Aldrich Co.
Gas Management Systems for GC
The information in this bulletin, in combination with the
instruction manual for your chromatograph, will help
you to use your gas management resources wisely and
obtain efficient performance from one, several, or
many gas chromatographs. Preceding information and
diagrams specific for installing gas delivery systems
for 1, 2-4, or 5-20 gas chromatographs is information
common to all installations: how to choose, clean, and
connect tubing, the comparative merits of gas cylin-
ders and gas generators, how to attain a suitable level
of gas purity, etc. We recommend you read this bulletin
and your instrument manual before you attempt to
install your GC(s).
Key Words:
? GC systems ? gas delivery ? gas management
? carrier gas
Foreword
One of the most important elements in ensuring optimum perfor-
mance from a gas chromatographic system is wise management
of the various gases required by the instrument. This bulletin will
assist analyst and lab manager alike in designing and installing
gas delivery systems for gas chromatographs. It encompasses all
aspects of gas management, from gas sources and gas purity to
safety considerations.
In designing a GC system numerous questions arise: What types
of gases will I need, and at what purity? Should I use compressed
gas cylinders or gas generators? Will this system meet my future
expansion needs? What are the safety and fire hazards, and how
do I handle them?
To approach these and many other issues in the most logical
manner, we have organized the information pertaining to system
design according to the complexity of the system. Simplest is the
single chromatograph installation with no consideration for ex-
pansion. More complicated is the installation of a bench of
(typically) 2-4 GCs. Most complicated is the system of several
benches or a lab of GCs, typically up to 20 units. Information and
diagrams specific to each of these situations are presented in the
latter part of the bulletin. Preceding the system-specific material
is information necessary for decision-making by all chromatogra-
phers: comparative merits of gas cylinders and gas generators,
how to decide on and attain suitable gas purity, how to choose,
clean, and connect tubing, testing for leaks, and other subjects
common to all systems.
Contents
Foreword
Basic Installation Concerns
Power Requirements 1
Gas Choices 2
Cylinders or Generators? 2
Gas Purity 4
Regulators and Associated Connectors 6
Tubing and Plumbing
Tubing Choices 7
Cleaning 8
Cutting – Reaming – Bending 9
Valves and Fittings 9
Making Connections 10
System Assembly 11
Finding and Eliminating Leaks 12
Purging 13
Purifier Connections 13
Installations
Single GC 13
2-4 GCs 14
5-20 GCs 24
Additional Resources 24
Basic Installation Concerns
Power Requirements
Whether you are planning a single GC, GC bench, or GC
laboratory system, we recommend you have a qualified electri-
cian review your power needs and recommend a suitable power
system. Be ready to provide an estimate of the total power
requirements for all GCs and associated equipment. A typical GC
consumes approximately 2100 watts and requires a 15-20 amp
dedicated, grounded outlet, to ensure it will not be affected by
transient signals from other sources (elevator or machinery
motors, vending machines, fluorescent lights, etc.). Add to this the
power needs of the integrator, plus peripheral equipment which
you anticipate using (autosampler, thermal desorber, pyrolizer,
etc.). The integrator or data system should be on the same outlet
or circuit as the GC from which it is acquiring data. This will help
prevent ground loop currents and reduce baseline noise. Equip-
ment requiring electric actuation, such as electric valve actuators,
should be on a separate line. Be sure that the outlets will be
located near the instruments, and will be in sufficient number to
meet current and future needs. Never use an extension cord of
any type or rating to connect a gas chromatograph.
It is important to have isolated and insulated grounding for these
instruments. In most plants, water lines and other sources of
grounding are used so heavily that they will not provide adequate
grounding. Maximum allowable line noise on a ground line is 3V
(rms), from 30Hz to 50Khz. We also recommend incorporating
surge protection in these lines.
SUPELCO
Bulletin 898
2
Table 1. Gases Used with Commonly Used Detectors
Detector Carrier Gas Fuel Gas Make-up Gas
ECD nitrogen, argon/ none nitrogen, argon/
5% methane 5% methane
ECD helium none argon/5% methane
ELCD, Hall® helium, hydrogen hydrogen none
FID helium, hydrogen, air + hydrogen nitrogen, helium,
nitrogen hydrogen
FPD nitrogen, helium air + hydrogen same as carrier gas
HID helium none helium
NPD helium, nitrogen, air + hydrogen helium
hydrogen
PID helium, hydrogen, none nitrogen, helium
nitrogen
TCD helium, hydrogen none same as carrier gas
Table 2. Gas Requirements of Gas Chromatograph
Systems with Flame-Type Detectors*
Gas Flow/Column (cc/min) Total (cc/min)
Dual packed column GC with 2 detectors
Carrier 20-60 40-120
Air (fuel) 350 700
Hydrogen (fuel) 30 60
Dual capillary GC with splitters and 2 detectors
Carrier 0.5-10 1-20
+ Split 100 200
+ Septum purge 4 8
Total 105-114 209-228
Make-up gas** 30 60
Air (fuel) 350 700
Hydrogen (fuel) 30 60
*Other detectors may not require fuel gases. See Table 1.
**Often, but not always, the same gas as the carrier.
Table 3. Hydrogen Output of
Packard Hydrogen Generators
Flow Max. Pressure
Model (cc/min) (psig)
9100 0-125 90
9200 0-250 90
9400 0-500 90
9800 0-1200 100
Gas Choices
The gases you will need for your chromatograph are a function of
the types of detectors you will use and the particulars of your
analyses. Table 1 lists typical GC detectors and the gases used
with each. The preference for one carrier gas versus another also
can differ from one analysis to another. A chromatograph equipped
with two typical flame-type detectors will require carrier, fuel
(oxidant) and, for some analyses, make-up gas, in the amounts
shown in Table 2. Consult your instruction manual for specific gas
requirements for your instrument.
List the types of detectors you anticipate using, and their carrier,
fuel, and makeup gas needs. You will need a separate line for each
gas. A general purpose system with several types of detectors
typically has five dedicated lines: helium, nitrogen, hydrogen, air,
and actuator (usually inexpensive compressed air), plus an auxil-
iary line. The auxiliary line anticipates a future need for a special gas,
such as argon or argon/methane or hydrogen/helium blends.
Do not use carrier, fuel, or makeup gas as an actuation gas.
Device actuation will temporarily disturb the gas supply to the GC
and affect its performance. Also, the quality of the gas used for
valve actuation is not demanding, so there is no need to use high
purity gas for this purpose. On the other hand, the actuation gas
must be oil- and particle-free, for long-term best performance from
the actuation equipment.
Cylinders or Generators?
Chromatographers traditionally have used compressed gas cylin-
ders, but today, primarily for safety and practicality, gas genera-
tors are becoming increasingly common. If you choose to use
cylinders or tanks, your gas supplier can help you determine the
sizes and numbers of cylinders you will need, and can help you
design the plumbing for your system. Your supplier can provide
cradles of 6 or 8 cylinders, already manifolded. A single line
connects the cradle to your house line. Depending on the size and
needs of your system, you also can use Dewars, bulk tanks, or
tube trailers as sources of compressed gases.
On the other hand, gas generators can greatly simplify plumbing
systems and eliminate the need for handling high pressure and/
or flammable materials. Because these compact units typically
can be located very near the instruments they serve, they elimi-
nate the need for long gas lines and cylinders mounted in
hallways. Compact, high purity, worry-free and safe generators of
nitrogen, air, and hydrogen are available. Hydrogen generators,
in particular, provide important safety advantages. Relative to
cylinders, the total amount of stored gas is small, and pressures
are much lower. This significantly reduces the risk of explosion.
Safety devices internal to most generators shut down the unit
when the pressure surges or suddenly drops. Maintenance time
spent on generators is less than that spent on changing cylinders.
Hydrogen Generator
Hydrogen generators electrolytically break water down into hy-
drogen and oxygen. The hydrogen is purified for chromatography
and the oxygen is vented. When used with a downstream water
trapping system, such as a molecular sieve trap and an OMI™
indicating trap in series, a hydrogen generator will provide GC-
quality hydrogen for both carrier gas and detector fuel use. New
models of hydrogen generators produce hydrogen at a purity of
99.99999+%, and internal purifiers in these models eliminate the
need for additional downstream purification.
To determine how many hydrogen generators you will need,
calculate expected flow needs based on the number of GCs and
the types of detectors and other equipment you will be using
(Tables 1 and 2 and your instrument manuals). Once you know
how much hydrogen you will need, you can determine which
model or models will meet that need (Table 3).
Hydrogen generators require deionized water of 500,000 ohm/cm
resistance, or greater, or a sodium hydroxide solution. At any
time, the total volume of gas in the unit is small, and the pressure
is low. Most units have a pressure relief valve, set for a pressure
slightly above the normal operating pressure. Other safety de-
vices within the generator also ensure pressures cannot exceed
the specified maximum, and shut down the unit if the pressure
suddenly drops. The emergency vent port on the back of any
hydrogen generator should be properly plumbed and safely
vented.
3SUPELCO
Bulletin 898
*Replace an oil-sealed
compressor with an
oilless unit to eliminate
the need for the particle
filter, oil-removing/
coalescing filter, and oil
vapor-removing filter.
Figure A. Filters and Traps Ensure High Quality Incoming Air for a Zero Air Generator (or for a Nitrogen Generator)
House Compressed Air*
Shutoff
Valve
Single-Stage
Regulator
Oil-Removing
(Coalescing)
Filter
Oil Vapor-
Removing
Filter Hydrocarbon Trap
GC
Shutoff
Valve
Pressure
Gauge
In-Line Filter
Shutoff
Valve
Zero Air
Generator
(or Nitrogen
Generator)
796-0066
Particle
Filter
Molecular Sieve 5A Drying TubeIn-Line Filter
Pressure
Gauge
Shutoff
Valve
Air Generator
An air generator is, in fact, a sophisticated air purifier. The source
of air typically is house compressed air or low-grade compressed
air cylinders. When properly installed (Figure A), a zero air
generator will provide air at a purity exceeding the quality de-
mands of your GC.
As with hydrogen, determine your air requirements from Tables
1 and 2 and your instrument manuals. One unit can provide ultra
high purity air to multiple detectors. Be sure to plan for extra
capacity, even if it means buying an extra unit – generators
should not be constantly operated at 100% of capacity.
Operating efficiency of a zero air generator is maximum if the
incoming compressed air contains less than 200ppm total hydro-
carbons and particles smaller than 7 microns. Compressed air
plumbing systems can contain rust, oils, and condensed liquids.
To remove oils, sulfur-containing compounds, and halocarbons
from the source air, install an oil-removing (coalescing) filter, a
vapor-removing filter, and a hydrocarbon trap before the genera-
tor (Figure A). If your plant air system does not sufficiently dry the
air, install a molecular sieve drying tube between the coalescing
filter and the generator inlet. A coalescing filter within the genera-
tor removes the last contaminants from the air.
A zero air generator operates best when supplied with com-
pressed air at 125psig or less and a flow of 2500cc/min or less.
Upstream from the hydrocarbon trap, install a single-stage pres-
sure regulator with a pressure gauge which will show operating
pressure to at least 125psig. We suggest a gauge which will show
pressures to 200psig.
Nitrogen Generator
There are two approaches to generating pure nitrogen. In one
approach, compressed air is passed across a semipermeable
membrane that allows nitrogen to pass and almost completely
bars other air components and contaminants. As the flow rate is
increased, however, oxygen breakthrough increases. The sec-
ond approach is a two-stage process. In the first stage, hydrocar-
bons in the inlet air supply are oxidized, producing carbon dioxide
and water. In the second stage, oxygen and the carbon dioxide
and water generated by hydrocarbon oxidation are adsorbed by
carbon molecular sieves (“pressure swing adsorption”). Both
methods work well – we recommend selecting a unit based on the
flow needs and purity requirements of the applications you intend
to use (remember to allow for future changes).
Passing the nitrogen leaving the generator through supplemental
purifiers can reduce the oxygen, carbon dioxide, and water levels
in the nitrogen even lower than the sub-parts-per-million levels
passed by the generator.
Calculate the nitrogen needs of your system, based on the
number of chromatographs and types of detectors you plan to use
(Tables 1 and 2 and your instrument manuals). Be sure to plan for
extra capacity – we do not recommend long-term operation of any
gas generator at full capacity. To obtain the best performance
from the nitrogen generator, remove water, dirt, rust, and oils from
the incoming compressed air in the same manner as for a zero air
generator (Figure A).
Be aware of the flow needs of the gas generator(s) you will be
using. With air generators there is an almost 1:1 ratio of incoming
gas flow to product gas – a little extra flow is suggested, but there
is almost no flow loss. With nitrogen generators, however, this is
not the case. Most of these units, independent of the purification
approach (semipermeable membrane or contaminant adsorp-
tion), require large quantities of input air to produce the desired
output flow. Typical air flow needs versus nitrogen delivered are
summarized in Table 4.
Compressors
Look critically at the source of your compressed air. Older facilities
typically have oil-sealed compressors. The longer they run to meet
your gas needs, the hotter they become. This leads to oil and
water vapors in the air stream. You can make either of two
choices: you can install a series of special filters to reduce the
hydrocarbon level in air leaving the compressor to less than
100ppm (particle filter, oil-removing/coalescing filter, and oil
vapor-removing filter, as shown in Figure A), or you can replace
the compressor with an oilless unit.
Most air compressors have built-in water vapor traps, but the heat
generated by the unit can cause significant amounts of water vapor
to still be present in the air produced. A water vapor trap down-
stream from the hydrocarbon trap will reduce the water content in
the air (Figure A). Depending on whether the compressor is oil-
sealed or oilless, the quantity of hydrocarbons will vary greatly.
Even an oilless compressor can allow hydrocarbon levels that
should cause concern. (The location of the air intake for the
compressor is very important in determining hydrocarbon levels).
SUPELCO
Bulletin 898
4
Figure B. Sealing Point in a CGA Connection
713-1109
Cylinder
Valve
Fitting CGA Connection
(CGA 590, for
purified air)
Seal made here, not at threads
Table 5. Acceptable Purity Levels for
Chromatography Grade Gases*
Impurity / Maximum Concentration
Total
Gas O2 H2O CO2 CO Hydrocarbons
Helium <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm
Nitrogen <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm
Air 20-22% <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm
Hydrogen <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm
Argon/ <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm
methane
*These limits are set to protect the column. Detector limits usually are less
demanding.
Table 4. Air Flow Need versus Output of
Nitrogen Generators
Nitrogen Air Consumption Nitrogen Delivered Stated Purity*
Generator (liters/minute) (liters/minute)
Air Products 75 1 99.99999%
Whatman
Models 75- 102 1.9 - 36.8 99.5 - 95%
Models 76- 42 0.5-2 99.99 - 99.9995%
*As flow is increased, purity is reduced.
Cylinder Safety
If you plan to use compressed gas cylinders, safety should be a
primary concern. A typical cylinder for analytical instruments has
a pressure of 2000-3000psig on delivery. A rupture at a cylinder
valve causes rapid depressurization and can cause serious injury
or structural damage to a lab.
Store cylinders in a secluded but easily accessible location. Avoid
humid places were rust can form on the caps or cylinders, and
locations heated by oven exhaust. Establish extra tie-down sites
near the cylinder in use, to hold the extra cylinder during the
changing process. In the lab, cylinder brackets should be bolted
to a wall or bench – brackets with screw clamps work satisfacto-
rily, but can become loose with long-term use. Wall-mounted
brackets with 1-3 cylinder capacity are available. A properly
secured cylinder cart is a safe alternative.
Always consider safety when changing cylinders or regulators. Do
not move cylinders unless you have a properly equipped cylinder
cart with chains to secure the cylinders in place. Never roll a
cylinder or move a cylinder with the cap off. Never change a
cylinder without safety equipment, including eye protection and
gloves. When changing cylinders, remove the expended cylinder
(label it “empty”), place it on the cart, and chain it in place, then
remove the new cylinder from the cart and install it. First secure
the new cylinder in place, then remove the cap. If the cap won’t
screw off, don’t try to force it. Do not place any object inside the
holes in the cap except a tool designed specifically for that
purpose. Return a cylinder with a seized cap to the manufacturer,
properly marked with the problem.
After removing the cap from the new cylinder, inspect the fitting
seat. Remove any dirt you observe – it could keep the fitting from
sealing properly, or it could be forced into your system. Carefully
screw the regulator onto the cylinder and tighten with the proper
sized wrench. Make sure the downstream pressure control knob
on the regulator is turned fully counterclockwise (valve closed).
Using two hands, open the cylinder valve while standing to the
side of cylinder. Never face the gauge(s) when opening a cylinder.
Bourdon tubes in pressure gauges can rupture with enough force
to cause serious injury. Slowly open the main regulator valve, then
slowly open the downstream pressure control knob and reestab-
lish the proper line pressure. Be sure to indicate on or near the
regulator the pressure the regulator should be reset to after a
cylinder change – you might not always be present when a
cylinder is emptied.
Use an electronic leak detector (never a liquid) to test for leaks
(see Finding and Eliminating Leaks: Testing for Leaks). If there
are no leaks, open the shutoff valve separating the cylinder and
regulator from the rest of the system. If you find leaks between the
cylinder and the system, close the cylinder valve (For the correct
procedure for relieving pressure in a two-stage regulator, refer to
Additional Comments on Regulators on page 7 of this bulletin.).
Unscrew the fitting and be sure there is no dirt on the fitting or
cylinder seat. If there is no dirt on the fitting and the leak persists
you may need a new fitting, or the seat of the fitting in the cylinder
valve may be damaged. If the fitting or the seat is damaged,using
Teflon® tape on the fitting will not work – the sealing point is at the
end of the fitting, not on the threads (Figure B).
Note: For detailed information on safe handling of cylinders and
regulators, refer to Safety Measures for Pressure Reducing
Regulators (Order from Air Products and Chemicals, Inc., 7201
Hamilton Boulevard, Allentown Pennsylvania 18195-1501 USA.)
When to Change a Cylinder
To be sure that the cylinder has not equilibrated with room air,
most gas suppliers request a minimum residual pressure of
25psig in a cylinder. If no pressure is present they must specially
clean and prepare the cylinder before re-pressurizing it. Also be
aware that as the pressure decreases in a cylinder the concentra-
tions of contaminants in the gas might increase, because they can
more easily pass from the liquid state to the gaseous state. This
is especially true of water – there will be a much higher concen-
tration of water in gas delivered at low pressure from a partially-
used cylinder than in gas from a new cylinder. For this reason, we
recommend changing carrier gas cylinders when the pressure
drops to 100-400psig.
Gas Purity
Gas purity in a GC system can be a confusing issue. In general,
chromatographers agree that oxygen, water, oils (hydrocarbons),
carbon monoxide, carbon dioxide, and halogens in gases supplied
to a chromatograph can cause baseline disturbances, ruin columns,
and/or damage detectors. Also, experiments confirm that gases
that meet the specifications in Table 5 will protect all GC equipment
discussed in this bulletin. Beyond these points, however, agree-
ment ends. Some chromatographers use only ultra-high purity
gases. Others, with equal experience, use lower grades of gas and
depend on in-line purification to provide the purity indicated in Table
5. Still others argue that gas purification usually is not needed and
fairly low grades of gases can be used routinely.
Several facts can bring a rationale to the gas purity arguments.
Water and oxygen damage a column by reacting with the phase.
Although exact temperatures at which this damage begins to
occur are not known, and probably differ among types of phases
and columns, reported damage consistently has been at tempera
5SUPELCO
Bulletin 898
Table 6. Recommended In-Line Gas Purifiers (Purifiers described on products pages)
Purifier Removes From Indicating?
Supelpure-HC Trap hydrocarbons all gases no
Molecular Sieve 5A water, heavy hydrocarbons air, hydrogen, nitrogen, no
helium
High Capacity Purifier oxygen, water helium, nitrogen yes
(heated purifier) do not use with hydrogen or air (pressure)
OMI oxygen, water, CO, CO2, argon, helium, nitrogen, yes
alcohols/phenols, sulfur- and hydrogen, (color)
halogen-containing compounds argon/methane, neon
Figure C. Face Seal Fitting
797-0086, 0085
Gland
GasketBody
GateKeeper
Female Nut Gland
Gasket Body
Weld Fitting
Female Nut
Weld Fitting
GateKeeper
tures of 140°C or above. Similarly, baseline disturbances due to
oxygen, CO, and CO2 are consistently reported at medium or
higher detector sensitivities. Therefore, if you intend to use low
temperatures and nondemanding detector sensitivities, you might
be able to use gases that do not meet the purity criteria in
Table 5. For all but a few selective detectors, however, you should
still be concerned about hydrocarbons in your gas, and use gas
which has very low hydrocarbon levels. Although some suppliers
of chromatography products are not concerned about hydrocar-
bons in gas streams, we strongly recommend using a hydrocar-
bon trap even if the gas is low in total hydrocarbons.
Gas Purifiers
Like the basic need for pure gases, the use of in-line gas purifiers
versus ultra high purity gases has been debated by analysts for
years. Because there are many sources of contaminants in
addition to the gas cylinder, we recommend using gas purifiers to
protect your instruments. Often, the greatest source of contami-
nants is the process of changing cylinders, which creates an
opportunity for room air to enter both line and cylinder. In-line
purifiers remove this surge of impurities and keep them from
entering the instrument. Unclean tubing can be a major source of
oils and other contaminants (see Tubing and Plumbing: Clean-
ing). Regulator diaphragms can be a source of hydrocarbons, and
oxygen can permeate through the diaphragm. Greases and/or
lubricants used in the body of a valve can be sources of hydrocar-
bons. Every fitting in the system potentially can allow room air, and
its associated contaminants, to leak into the system. Even a
system that is initially leak-free can develop leaks over time, due
to expansion and contraction of tubing and fittings with the
changing temperatures in the lab. Indicating (color changing) in-
line purifiers, available for oxygen, water, hydrocarbons, and
other contaminants, give visual warning that contamination is
present. In some purifiers, a pressure drop develops between
inlet and outlet as the purifier’s capacity is reached, and acts as
the signal for changing the purifier. This means installing pressure
gauges at each end of the purifier and routinely monitoring the
pressure. Typically, the purifier should be changed when the
pressure drop reaches 10-15psig.
For ultra-demanding applications requiring the highest possible
gas purity, there are special purifiers and connectors. The Aeronox
GateKeeper purifier reduces contaminants to the sub-parts-per-
billion level. A combination of special design factors allows the
purifier to reach these levels. The catalyst, a nickel material,
reacts with a variety of contaminating materials and permanently
removes them. Specially electropolished inner surfaces and
special endfittings – face seal fittings (Figure C) – then maintain
the purity level. (Standard compression fittings are very good, but
cannot completely eliminate trace leaks that allow ppb levels of
contaminants.) Few chromatographers, however, need this chal-
lenging level of gas purity.
Carrier gas purification should start with large capacity (bulk)
purifiers for trapping hydrocarbons, water, and oxygen, in that
order, in the main gas lines. If the order is changed the lifespans
of some of these devices could be shortened. Smaller capacity
purifiers and in-line filters should be installed in each carrier gas
and make-up gas branch line, as close to the GC as possible
(Table 6). The OMI™ purifier provides final purification of carrier
or make-up gas. Its capacity is smaller than that of bulk purifiers,but
it will provide many months of operation if lines are leak-tight and
properly maintained. The color indicator in an OMI purifier changes
from black to brown as the material (Nanochem® resin) is ex-
pended. For more information on the OMI purifier request Bulletin
848. We also recommend using a hydrocarbon trap in the fuel gas
line.
Additional Comments on Gas Purifiers
The color change line visible through the wall of a color-indicating
purifier is not the most forward point at which the purifier material
is expended. There is a tunneling or funneling aspect to the
purification process – the coe of the purifier is expended before
the outside edges. Therefore, you should change the purifier
when the color change is about 75% along the tube – the front of
the cone of expended material will be much closer to the outlet end
of the tube. For convenience, we mark the “time to change” point
on the body of the OMI tube (Figure D).
SUPELCO
Bulletin 898
6
Figure E. Gas Flowpaths in Single-Stage and Two-Stage Regulators
Outlet
Range Spring
Filter
Captured Gas Port
Vent Diaphragm
Inlet
Single-Stage
Regulator
Adjusting Knob
Two-Stage
Regulator
Inlet Outlet
1st Stage
Valve
Stem
Captured
Gas Port
Captured
Gas Port
2nd Stage
Range Spring
1st Stage
Range Spring
2nd Stage
Valve Stem
Intermediate
Pressure
Low Pressure
Adjusting Knob
Valve
Stem
796-0054, 0055
High Pressure
Low Pressure
Time-to-Change Indication
996-0054
Figure D. Partially Used OMI Purifier with
“Time-To-Change” Marking
? Gas generators develop much lower gas pressures than the pressures delivered
from cylinders. A single stage regulator is suitable for regulating gas flows from
these devices.
Not all purifiers on the market are adequate for use in carrier gas
lines. Purifiers constructed from plastic tubing tend to allow water
and oxygen to permeate into the gas line. Most water-removing
purifiers using Drierite® and similar color indicators do not com-
pletely remove water. Also avoid purifiers which have O-ring
seals; they typically leak, especially on re-tightening or on the
second installation. We recommend the purifiers in Table 6 for
removing the gases indicated to levels acceptable in carrier gas.
Purifiers constructed of plastic, or filled with low-efficiency
adsorbents, are adequate for air lines.
In-Line Filters
To protect needle valves, regulators, flow controllers, and other de-
vices, each gas line should contain a filter capable of removing particles
7-10 microns in diameter (see products pages of this bulletin). The filter
in a two-stage regulator will not trap particles this fine.
Regulators and Associated Connectors
Two types of regulators are used in a well-designed gas supply
system. At each gas cylinder is a two-stage regulator.? The first
stage reduces the pressure of gas coming from the cylinder to 300-
500psig, then the second stage reduces the pressure to
the pressure desired in the main line (Figure E). If you use a single-
stage regulator at the cylinder, you must constantly adjust the main
line pressure as the pressure in the cylinder is reduced. With a
single-stage regulator, downstream line pressure will increase at a
rate of 0.65psig per 100psig decrease in cylinder pressure. This
change in downstream pressure may be unacceptable. (Note: two
pressure gauges do not always denote a two-stage regulator.
Some single-stage regulators have an inlet and outlet gauge.)
In multiple-unit GC systems, the branch line to each chromato-
graph should include a single-stage regulator, to step down the
pressure in the line to that required by the instrument.
There is another reason why both two-stage and single-stage
regulators are used in a GC system. To ensure effective operation,
you must maintain at least a 10-15psig pressure differential across
all flow and pressure-controlling devices (Figure F). Pressure in the
mainline can change because of new demands, because the
cylinder pressure output is not properly reset when a cylinder is
changed, because the system has long plumbing lines (pressure
will be lowest at the most remote instruments), or because pressure
will vary during a temperature programmed analysis. To ensure
that you maintain a 10-15psig pressure differential, you must know
the pressure of the gas as it enters the GC and at the head pressure
gauge on the instrument. A system with a two-stage regulator at the
cylinder and a single stage in-line regulator at each chromatograph
provides this information. Without a single-stage regulator just
before each instrument, changes in the main line pressure will
affect the operation of the individual GCs.
Not all commercially available regulators are suitable for use with
GC carrier gases. The critical component is the diaphragm. The
Buna-N or neoprene diaphragms in most regulators off-gas
contaminants and are permeable to water and oxygen (see Scott
Specialty Gases laboratory report No. E-R83-1, request from
Scott Specialty Gases, Plumsteadville, PA, USA). A regulator with
a stainless steel diaphragm eliminates these problems. On the
other hand, regulators constructed entirely of stainless steel,
intended for use with corrosive gases, are very expensive and are
not needed for GC applications.
7SUPELCO
Bulletin 898
Figure F. To Ensure Effective Operation, Maintain
At Least a 10-15psig Pressure Differential Across
All Flow and Pressure-Controlling Devices
Line Regulator*
60psi 60psi 60psi
90psi 90psi90psi
75psiGC 75psiGC75psiGC
796-0067
* Simplified diagram, does not show shut-off valves above and below
pressure regulator (see Figure X).
Table 7. Regulator Fittings
CGA Connections (all 0.965")
Outlet Nipple
Gas Description Shape Nut Thread*
Air (purified) 590-14LH bullet male left
Argon 580-14RH bullet male right
Argon/Methane 350-14LH round female left
Helium 580-14RH bullet male right
Hydrogen 350-14LH round female left
Nitrogen 580-14RH bullet male right
Other Connections
Gas DIN 477 BS 341 Japan
Air (purified) 13 3 —
Argon 6 3 W22-14 - right
Argon/Methane 1 4 —
Helium 6 3 W20.9-14 - left
Hydrogen 1 4 W22-14 - left
Methane 1 — —
Nitrogen 6 3 W22-14 - right
*There is always a groove in a nut with left-handed threads (see Figure B).
Regulators used in an air line can be fitted with a neoprene
diaphragm. These regulators will reduce installation costs without
sacrificing the integrity of the system.
Be sure to read the product specifications before you buy a
regulator. A regulator with two pressure gauges is not necessarily
a two-stage regulator.
Additional Comments on Regulators
Modern gas chromatographs are factory set to operate at column
head pressures of up to approximately 60psig. In some GCs, very
long columns (e.g., 100+ meter capillary columns) can require
column head pressures up to 90psig. Typically, the line pressure
should be 15-20psig higher than the inlet pressure to the GC and,
as discussed above, for pressure regulators and flow controllers
to work correctly there must be a 10-15psig difference between
the input pressure and the maximum output pressure the devices
will be expected to deliver. Consider these values in deciding what
pressures to use in your system – higher pressures may require
you to change some internal pressure gauges in your GC.
Never remove a two-stage regulator from a gas line with a high
pressure isolated in the first stage – the sudden release of
pressure could rupture the diaphragm, ruin diaphragms in down-
stream regulators, and/or create gaps in a packed column (the
packing could even be forced out of the column). Always depres-
surize a two-stage regulator through the second stage. If your
system has a single-cylinder gas supply, or a gas generator, the
first step is to turn off the GC oven and let the column cool. In a
two-cylinder system, transfer flow to the second cylinder. Next,
close the first stage (cylinder side) valve on the regulator to be
removed from service, while leaving open the shutoff valve after
the regulator. This will allow the gas remaining in the regulator to
pass through the regulator. Vent the pressure through the system
(be sure the column is cold), through a vent installed in the gas
line, or through the vent on the regulator itself (some models).
Finally, close the downstream pressure control valve and remove
the regulator.
Whenever you change cylinders or regulators, be sure to protect
the columns in the chromatograph. Before you disrupt the gas
flow, either switch to a second source of gas or, if you are
disrupting carrier gas flow, turn off the oven(s) and cool the
column(s) supplied by the gas.
Always use a regulator rated for your intended application. Never
switch CGA, or other, fittings to use a regulator for a purpose for
which it was not intended (e.g., do not refit any regulator for
oxygen delivery). Table 7 lists the proper CGA fittings for each
type of gas used for GC. Never switch gauges or inlet fittings and
never change gas service. Never close a regulator body in a vise
to remove a fitting – this almost certainly will break the diaphragm
seal and cause the diaphragm to leak. Never lubricate a regulator
or use pipe sealants.
The regulators we offer supply a maximum of 100psig and have
gauges that read to 200psig in 5psig intervals. Regulators that can
provide much higher pressures are available, but we do not
recommend these for GC use. It is difficult to regulate a pressure
to within a few psig on a 2000psig regulator with gauge gradations
in 20psig intervals.
Be sure to indicate, on or near each regulator in your system, the
pressure the regulator should be reset to after a cylinder change
– you might not always be present when a cylinder is emptied and
replaced.
The American Society for Testing and Materials (ASTM) sum-
mary book on chromatography, and related ASTM publications,
provide “lab-tested” guidance to the practicing chromatographer
(order from American Society for Testing and Materials, 100 Bar
Harbor Drive, West Conshohocken, Pennsylvania 19428-2959
USA).
Tubing and Plumbing
Tubing Choices
Many types of tubing are available for supplying air, helium,
nitrogen, argon, and gas mixtures for gas chromatography. In
practice, however, only copper and stainless steel are viable
alternatives. Table 8 lists the specifications for tubing suitable for
use in gas chromatography. Before use, this tubing must be
cleaned to remove traces of oil and dirt. Regular grade metal
tubing offered in chromatography catalogs usually is adequately
cleaned for plumbing purposes. Column-quality tubing receives
additional acid and base cleaning and is chromatographically
tested for active sites. This extra treatment is not needed for gas
lines. Tubing obtained from building supply houses or hardware
stores is not clean enough for use – our chemists have seen oil
SUPELCO
Bulletin 898
8
Figure G. Tubing Crimped and Folded
to Keep Dirt Out
796-0074
Table 9. Recommended Lengths of GC Gas Lines
Outside Maximum Length
Material Diameter (feet)
Copper* 1/8** 6
1/4 20
1/2 100
Stainless Steel 1/16 2
1/8 10
1/4 20
1/2 100
*Do not use with hydrogen.
**Not recommended due to brittleness.
Table 8. Tubing and Tubing Preparation
for Gas Chromatography
Tubing Type For Columns For Plumbing
Stainless Steel* premium grade 304 regular grade 304
Copper** highly cleaned cleaned
Aluminum highly cleaned not recommended
Nickel SP-Alloy (T-1) not recommended
Teflon® TFE or FEP not recommended
Tygon® not recommended air lines only
*Conforms to Schedule 40 ASTM A213, RB80
**Type K hard tempered
dripping from the vent ports of TCD detectors when non-chro-
matographic grade tubing was used, uncleaned, to plumb a GC
system. Similarly, the dirt in commercial tubing can clog the frits
in flow controllers and other fine metering valves, and ruin these
devices. To be sure of the quality of the tubing used in your lab,
discuss the Tubing and Plumbing section of this bulletin with the
suppliers and fabricators of your gas supply system.
Stainless Steel Tubing — Strong and reusable, stainless steel
tubing is always the best choice – and the most costly – for a GC
system. For hydrogen, mill-finished, oxygen-cleaned 304 or 316
grade stainless steel tubing (never copper) should always be
used. Care should still be taken to clean this material. For special
applications where pristine conditions are needed, such as with
helium ionization detectors, 304L electropolished (EP) stainless
steel with vacuum coupled replaceable (VCR) connections and
orbital weld joints is the best choice. An electropolished surface
significantly reduces water and contaminant capture. It can be
very expensive, but for critical applications it is worth the added
expense.
Copper Tubing — Due to its lower cost, copper tubing is the most
commonly used plumbing material in GC systems. Copper should
not be used with hydrogen gas, nor where the gas line might be
flexed. With time copper tubing of any diameter work-hardens and
is very easily broken during flexing. Because copper tubing has a
much smaller inside diameter than stainless steel tubing (1.65mm
versus 2.1mm), only very short lengths of 1/8” copper tubing
should be used (Table 9). Long lengths lead to high back pres-
sures. (Even stainless steel gas lines should be as short as
possible.) 1/4" OD copper tubing is by far the most common diameter
of copper tubing used – it tends to be stronger than 1/8”tubing, but with
flexing it can still break. 1/2” OD copper tubing typically is
inflexible. Consequently, Swagelok® or soldered fittings are needed
for all direction changes and connections in 1/2” copper lines.
Additional Comments on Tubing
Never use cast iron or black steel pipe to supply gases to
chromatographs. Over time these materials will form rust which
will travel through your system, ruining valves, regulators and
other components.
Soft and easily kinked when new, aluminum tubing, like copper
tubing, becomes brittle over time. Because aluminum tubing
offers no particular advantages relative to copper, and has a
higher degree of the same disadvantages, we do not recommend
using aluminum tubing to plumb a GC system.
Teflon and nylon tubing are acceptable for air and actuation lines,
but permeability to water and oxygen precludes the use of these
or other polymeric materials for most GC plumbing needs, includ-
ing carrier gas and make-up gas lines. Hydrocarbons from some
polymeric tubing can appear as impurities in the system. In a well-
designed plumbing system, a regulator can fail and release full
cylinder pressure into a line, and the line will withstand the
pressure. Polymer tubing will not pass this test.
Cleaning
Dirt and oil are present in all tubing as a result of the manufacturing
process. Only if you buy cleaned tubing from a chromatography
supplies dealer will you receive tubing immediately ready for
chromatography. Even then the tubing must be capped to keep
dust and dirt out during shipment and the system assembly
process. During cutting and assembly processes, metal frag-
ments and dirt can get into the tubing. It is best to clean the tubing,
assemble it, blow it out, and purge it.
The first concern is the removal of dirt and oil used during the
manufacturing process. If you have any intention of using an
electron capture detector, at any time, do not clean the tubing with
chlorinated solvents. Using a nonpolar solvent such as n-hexane,
flush the tubing until the solvent flowing out of the line is clear.
Allow sufficient time for the solvent to dissolve materials in the line.
Rinse the tubing with water to flush the hexane and absorb any
free acidic or basic material. Next, flush with methanol, to remove
traces of hydrocarbons and the water remaining in the tubing.
Using clean nitrogen (not compressed air, which always contains
some oils), attempt to remove all traces of methanol. Coiled tubing
can be put in a large oven and heated to 110°C during the nitrogen
purge.
Clean tubing should be capped in some fashion to keep dirt out.
If caps are not available, flatten the ends of the tubing and fold
each flattened end back on itself (Figure G).
9SUPELCO
Bulletin 898
Figure H. Uncoil Tubing without Creating Bends
996-0056
Figure I. Trim and Bevel a Newly Cut Tubing End
996-0057
Figure J. Conventional (left) and Hydrogen
Pressure Relief Valves
796-0075
Beveled
edge
Cutting – Reaming – Bending
To avoid creating kinks or flat spots as you uncoil the tubing, hold
the coil of tubing perpendicular to a table or the floor. Hold the end
of the tubing with one hand and roll the coil away from you with
your other hand (Figure H).
The preferred tool for cutting copper or stainless steel tubing is a
device which presses a cutting wheel against the tubing while the
device is turned repeatedly around the tubing (see the products
pages of this bulletin) This tool makes a very clean, truly perpen-
dicular cut which allows the tubing to fit squarely into a fitting.
Some cutting machines also work very well, especially for cutting
1/16" tubing. In contrast, general-purpose tubing cutters distort
the end of the tubing and hand-held saws often leave ragged and
angled cuts and excessive metal filings in the tubing.
Cutting should always be followed by reaming. When cutting any
tubing, but especially when cutting copper, a soft metal, the metal
typically intrudes inward and reduces the ID of the tubing, some-
times almost completely closing it. Special care must be given to
re-opening the tubing to its original inside diameter. Use a tool
such as Catalog No. 20389 (see the products pages) to carefully
cut away excess metal and slightly bevel the inside edge of the
tubing (Figure I). Remember to clean the metal filings out of the
tubing, or they will be pushed into the nearest valve, flow control-
ler, or pressure regulator, where they could cause damage. Direct
a steam of clean, dry nitrogen gas through the tubing to remove the
filings. Do not use air from a compressor; it might contain oils.
Bend tubing very carefully, taking care to not reduce the inside
diameter or create flat spots. Use tools designed for this purpose,
such as Catalog Nos. 20422-U, 20424-U, and 20857 (see the
products pages). If the tubing at a bend is visibly flat, discard it. If
you need a very sharp bend or there isn’t room for a bend, use an
elbow fitting.
Flexible Hoses — Most tubing is not designed for continual flexing.
This creates a problem for attaching a regulator to a cylinder or to
a gas line. The solution is a flexible metal hose (see the products
pages) – a 30" length of corrugated (bellows) stainless steel tubing
reinforced with stainless steel braids, with additional casing on the
outside, and fitted with Swagelok, pipe thread (male) or CGA
connectors. The hose can be used to connect a cylinder to a fixed,
wall-mounted regulator or to connect a regulator mounted on a
cylinder to a gas line that is secured to a bench or wall. The hose
should be rated to 3000psig and for the gas you will be using. When
the regulator is removed from the cylinder it should be properly
supported – do not suspend it by the hose.
Valves and Fittings
In addition to the tubing, all other system components – joints,
valves, relief valves, flash arrestor etc. – must be compatible with
anticipated operating pressures and temperatures. Trace con-
taminants usually come from O-rings, washers, elastomers, and
plasticizers sometimes used in valves or other devices. Avoid this
problem by eliminating elastomer valve seats and using metal-to-
metal seals for all joints and seals (Military Specification grades of
Teflon, T-27730A may be acceptable). Greased fittings and
soldered lead joints should not be used because of potential
contamination from organic greases or acid solder flux. When
using copper tubing, Swagelok fittings or well fabricated brazed
joints usually provide leak tight connections. 1/4-turn or 1/2-turn
bellows- or diaphragm-type valves assure the best positive shutoff
of gas flow (see the products pages).
Pressure Gauges — Pressure gauges should be selected by
pressure range. They should exceed the pressure you anticipate
using, but not greatly. It is very difficult to read 10- or 20-psig
increments on a 2000psig gauge. In most GC systems, none of
the gases being used will ever exceed 100psig. Pressure gauges
have pipe threads. They should be attached using Teflon tape.
Never use pipe sealant.
SUPELCO
Bulletin 898
10
Figure K. Safely Designed Hydrogen Line (Carrier Gas)
Pressure
Relief
Device
Flashback
Arrestor
Vented to Fume
Hood (or other
approved outlet)
Shutoff
Valves
Hydrogen
Cylinder
Hydrocarbon Trap
Two-Stage
Pressure Regulator
Shutoff Valve
Pressure
Gauge
In-Line Filter OMI Tube
Molecular Sieve 5A Drying Tube
GC
796-0068
Pressure Relief Devices — A pressure relief device is required
with any flammable gas (e.g., hydrogen), whether delivered from a
generator or a cylinder. A hydrogen pressure relief valve is different
from most pressure relief valves in that it is designed to accept
fittings which allow additional plumbing and proper venting (Figure
J). Most hydrogen generators will have an internal device which
should be properly plumbed to a safe vent. If you cannot confirm that
your generator has such a device, install one just downstream of the
generator, in conjunction with a flash arrestor (Figure K).
The pressure relief device in a hydrogen line should always be
safely vented. Mixtures of 4% or more hydrogen in air are explosive.
Do not allow these concentrations to form in the lab. Vent hydrogen
to a fume hood or other conduit leading out of the building. Check
with your safety department to determine the proper venting
procedure for your site.
We recommend you install a pressure relief valve that releases
pressures above 2000psig on each main gas line, to protect
downstream equipment from high pressure failures. The best
location for the device is after the cylinder regulator shutoff valve
(Figure K). Pressure relief valves on non-flammable gas lines need
not be vented to a hood, but be sure to direct the vents downward
(away from operator).
Dry Flashback Arrestor — In the event of a hydrogen flashback,
a dry flashback arrestor diverts the flame into a 3-foot (1m) length
of tubing, where the flame is extinguished and the heat is absorbed.
The shock wave preceding the flashback closes and locks the
arrestor’s shutoff valve, eliminating continued gas feed. Install the
dry flashback arrestor after the shutoff valve and pressure relief
device for the cylinder regulator (Figure K). Many hydrogen genera-
tors incorporate a flashback arrestor. If your generator does not,
install one just downstream of the generator. Use only devices
which meet Occupational Health and Safety Administration (OSHA)
and National Fire Protection Agency (NFPA) codes, or overseas
equivalents, and are Factory Mutual approved. Devices larger than
the one shown on the products pages of this bulletin may be needed
for large installations. They are available commercially. Dry flash-
back arrestors are reusable and can be reset, but be sure to
determine and eliminate the cause of the flashback before resetting
the arrestor.
In contrast, wet flashback arrestors, which incorporate ethylene
glycol, should not be used with chromatographic systems. Al-
though ethylene glycol is only weakly volatile, ethylene glycol vapor
could be released into the gas system. This contamination will
cause unstable baselines and high background signals.
Making Connections
Installation of all the lines, regulators, valves and other associated
hardware needed in a GC system requires an assortment of tube,
threaded pipe and, perhaps, soldered connections. When tube
connections are required, always use highest quality fittings. We
recommend using Swagelok fittings wherever possible. Threaded
pipe connections should be sealed only with instrument grade
Teflon tape (Catalog No. 20808-U). Pipe sealant (pipe dope) or
other chemicals, and some lower grade Teflon tapes, contain
organics which could bleed into the gas stream. Roll one layer of
the tape onto the threads counter to the direction of the threading (i.e.,
counterclockwise) and tighten the tape. Thread the two parts together and
tighten. Do not overtighten.
When soldered connections are needed, the brazing alloy should
be flat stick silver solder containing 15% silver. Use MAPP® (methyl
acetylene propadiene), rather than acetylene, when soldering with
this high melting point solder. Use no flux. Flux will cause interfer-
ence with electron capture detectors, and possibly with other
detectors and some columns.
Assembling a Swagelok Tube Fitting
Before assembling a nut and ferrule on the tubing inspect the tubing
to be sure the surface is smooth and free of longitudinal scratches,
and the cut end is deburred. If the tubing is acceptable, slide on the
nut with the open side facing the end of the tubing. Next slide on the
back ferrule with the wider part facing the nut. Then slide on the front
ferrule with the small end of the cone facing the end of the tubing
(Figure L). Push the assembly about 1" (2-3cm) onto the tubing.
The ferrules and nut should slide onto the tubing easily and rotate
freely. Insert the tubing into the fitting – it should fit easily. Hand-
tighten the nut/ferrule assembly onto the fitting. Then, using two
wrenches, tighten the assembly. We do not recommend relying on
torque measurements, due to differences in tubing wall thickness
and materials of construction.
11SUPELCO
Bulletin 898
Figure L. Assembling a Swagelok Fitting
Back Ferrule
Tubing
796-0076
Nut
Front Ferrule
Front Ferrule
Back Ferrule
Tubing
Nut
M-2: Overtightened Fitting
Front Ferrule
(Note concave
profile)
Nut
Back Ferrule
Tubing
796-0077, 0078
Figure M. Ferrule Profile Reveals
Correct/Incorrect Tightening
M-1: Properly Seated Fitting
Instead, monitor the number of turns you make on the nut. If the
parts are clean and properly assembled, 3/4 turn on 1/16” or
1/8” tubing, or 1¼ turns on 1/4" tubing, should seal the fitting. A
properly tightened Swagelok ferrule system looks like Figure
M-1. Notice that a properly seated front ferrule will be forced
slightly into the tubing. Always use two wrenches when tightening
fittings, one to hold the fitting in place and the other to tighten the
nut/ferrule assembly. A tee wrench is very useful for tightening
tees. A hydraulic swaging unit might be required toswage 1/2” or
larger fittings. If the fitting does not seal properly, additional
tightening seldom provides a leak free seal. Disconnect the nut
and examine the inner surfaces of the fitting, ferrules, and tubing
for dirt or scratches. If necessary, replace defective components.
Figure M-2 shows the effects of overtightening a fitting. Notice the
concave front ferrule. The shoulder on the distorted ferrule is
typical of overtightened ferrules. The cross-sectional profile of a
good ferrule is a straight edge from the tubing to the back edge of
the ferrule. Never use sealing compounds on the outside of
fittings to stop leaks.
For detailed instructions on swaging and other plumbing needs,
refer to the Swagelok Manual (22339).
There should be no need to disassemble and inspect a Swagelok
fitting if it passes your leak testing procedure, but in some facilities
additional testing is required to ensure that a fitting has been
sufficiently tightened. The preferred test involves using gap
inspection gauges (Catalog Nos. 21984-U, 21985-U and 25822).
A gap inspection gauge has a thick end and a thin end. Attempt
to insert the thick end of the gauge in the gap between the nut and
body of the tightened tube fitting. If the thick end will fit, the fitting
nut has not been sufficiently tightened. If the thick end will not fit
the minimum requirement for tightening has been met.
Properly installed Swagelok fittings can be disconnected and re-
connected many times. To re-connect a fitting simply hand-
tighten the nut, then slightly tighten it with a wrench. It should take
little additional pressure to tighten the fitting, because you are
simply making a metal to metal seal between the ferrules and the
body of the fitting, not re-seating the ferrules onto the tubing.
Always confirm the reconnection is leak-free, using an electronic
leak-detecting device.
Additional Comments on Connections
Never mix tube fitting components of different brands. Although
products from different manufacturers appear to be interchange-
able, they are not. Nuts, ferrules, and bodies will have different
angles and depth specifications. Two-piece ferrules and one-
piece ferrules have different mechanical sealing functions be-
tween the tubing and the fitting body. Decide on a fitting manufac-
turer and stay with the decision throughout your plumbing system.
For information on installing a GC column into the system, request
our free Bulletin 741, The Supelco Guide to Leak Free Connec-
tions: Ferrules and Fittings for Packed and Capillary GC.
System Assembly
Your plumbing assembly should follow either the single GC, 2-4
GC, or 5-20 GC system described later in this bulletin. Figures in
these sections show our recommendations for the various types
of valves, regulators, and other devices for each system. We also
recommend that you read the Swagelok Tube Fitter’s Manual
(Catalog No. 22339), especially chapter 3, Tubing and Tube
Fitting Handling and Installation. The manual offers many tips and
helpful directions which go beyond the detail in this bulletin.
Securing Fittings and Tubing
Because it is not desirable to flex the tubing when opening or
closing a valve, valves and gas lines should be securely fastened
to benches or walls. Many types of fasteners are available for
tubing, and brackets are available for most valves (see the
products pages). Tubing should be fastened down every 4-6 feet.
Though most of the plumbing in your system should be secured to
a bench or wall, there should be some flexibility at the point of
connection to the GC. It is a good idea to roll about 3 feet of the gas
line between the shutoff valve and the GC into a coil 4-6 inches in
diameter (Figure N). This will allow some lateral and front-to-back
movement of the instrument when it is being serviced.
Plumbing Two Gases Together
Some analysts frequently switch carrier gases as the application
for an instrument changes. Never attempt to plumb two carrier
gases through the same line, through tees, valves, or other
arrangements. Even with shutoff valves, flow-check valves, and
other devices which are supposed to guarantee that the two gases
never mix, in time they do mix, usually through human error. It
takes little time to disconnect one line and attach another, and this
is the best approach. If time is critical, you can use quick connect
fittings for this purpose. If you use quick connect fittings, however,
we strongly suggest you routinely test for leaks, and include
oxygen and water vapor traps downstream.
SUPELCO
Bulletin 898
12
Figure O. Leak Checking a Valve
with a GOW-MAC Leak Detector
910-0076
Check all
potential leak
points
Vent
Hole
996-0055
Vent
Hole
Figure P. Be Sure to Leak Check All Regulators
Figure N. Coiled Tubing Allows
Flexibility in Installation of a GC
9960058
Coding Gas Lines
It is important to know what gas a valve will deliver when you open
it, but when all the gas lines are in place it can be difficult to discern
what line contains what gas. In Supelco laboratories each line is
color coded (i.e., painted) and labeled so that analysts and repair
technicians can quickly determine what gas each line contains.
You can buy colored sleeves, tags, or other types of labels, or
simply paint the lines in different colors.
Finding and Eliminating Leaks
Equipment Alternatives
Testing a system for leaks is often thought of as being very
difficult. In truth, the initial pressure test of the entire system is very
simple. If the system passes this initial test it is ready to use.
Finding the leaks, if they exist, can be more difficult. Often a
cylinder of oil-free air is used to sequentially check each of the gas
lines in a new system for leaks – it is costly to use high purity
gases, and it is not safe to leak test with hydrogen. On the other
hand, helium or nitrogen will allow electronic leak testing if part of
the system fails the pressure test. We recommend using high-
purity helium, or a lower grade of helium passed through traps that
will remove hydrocarbons, oxygen, water, and particles (see the
Installations section of this bulletin). Helium is the easiest gas to
detect with an electronic leak detector.
Pressure testing reveals the presence of leaks, but does not show
where they are located. You must find leaks by using either a liquid
or an electronic leak detector. We recommend that all leak testing
be done with an electronic leak detecting device, not with liquids
of any kind. Just like a kitchen sink aspirator will draw a vacuum
on a small side line while water is running through the main line,
a leak –small or large – will draw in gas or liquids as it allows gas
to leak out. If there is a leak in a line, any liquid leak detector could
be siphoned into your system and could reduce sensitivity or
cause a drifting baseline. To avoid any chance of contamination,
we strongly recommend using an electronic leak detector.
GOW-MAC electronic leak detectors are simple to use. Simply set
the read-out to zero as only air is being drawn into the unit, then
place the probe at the site to be tested and sample the air around
the site (Figure O). The detector, a form of TCD, senses the
thermal conductivity of the gas in the detector cell. If the detector
senses gas mixtures other than normal air, the needle on the
gauge will be deflected, indicating a leak. The detector is very
sensitive for helium and hydrogen. Although less sensitive for
nitrogen (air is 80% nitrogen, so the differences in thermal
conductivity are small), its nitrogen sensing capability is as good
as that of liquid leak detectors. Obviously, you cannot use this
device to test air lines for leaks.
Testing for Leaks
Caution: Always wear eye protection and gloves when opening
or closing cylinders. Do not stand in front of the gauges. Bourdon
tubes in pressure gauges can rupture with enough force to cause
serious injury.
Note: Bypass or remove purifiers during leak testing.
Open all valves in the line (and in each branch line, in a multiple-GC
system), but close the last shutoff valve just before the GC(s).
Pressurize the system to 100psig. After a few minutes, when the
pressure is stable, close the shutoff valve immediately downstream
from the cylinder regulator. You may see an initial pressure drop of
a few pounds (it might be necessary to install a pressure gauge after
the shutoff valve for this test). If the system then
13SUPELCO
Bulletin 898
maintains pressure for 1/2 hour you have no leaks of any significance.
If the pressure continues to drop, you will need to search for leaks
(typically it will drop very rapidly if leaks exist). In a multiple-GC
system, shutoff valves at each branch and just before each GC
enable you to isolate and test sections of the system. This is quicker
than testing every fitting, regulator, and valve.
If your system fails the pressure test, and you have been pressure
testing with air, vent the air in the section(s) involved and re-
pressurize with helium. Using an electronic leak detecting device,
systematically isolate and test each section, starting at one end of
the system and working back to the cylinder. If you find a leak, seal
it, then pressure test again before proceeding. Test each line
connection, each valve knob connection, the vent holes of regula-
tors, or anywhere you feel a leak could develop. Don’t forget to test
connections inside the chromatograph.
Most regulators have a small vent hole on the spring side of the
diaphragm (Figure P). If gas is leaking from this site, the diaphragm
or an internal seal is bad. If the leak is from the pressure relief port,
that part of the regulator may be bad. If the leak is internal to the
regulator – a leaking diaphragm, internal seal, or pressure relief
vent – replace the regulator. Never attempt to repair or replace any
parts other than the CGA fitting. Faulty regulators should be sent to
people who are properly trained in regulator maintenance.
If the leak is between the CGA fitting and the regulator the fitting may
simply be loose and need tightening. Most CGA fittings have flat
surfaces for tightening. Don’t overtighten. If the leak persists,
unscrew the CGA fitting and be sure there is no dirt on the fitting or
cylinder seat. If dirt is not the problem and the leak persists you may
need a new CGA fitting, or the seat of the fitting in the cylinder valve
may be damaged. If the fitting or the seat is damaged, using Teflon
tape on the fitting will not work – the sealing point is at the end of the
fitting, not on the threads (Figure B). After you find and seal all leaks
it is time to fill the lines with the correct gases.
Purging
Once you have determined that the system is leak free, you are
ready to purge the lines and replace the air, nitrogen, or helium test
gas in each line with the gas for which the line is intended. The
procedure to follow depends on the gas line you are purging. For air
and hydrogen being used as detector fuel, use the simple purging
procedure for fuel gases described below. Carrier gas lines and
make-up gas lines require a more extensive purge to ensure the
desired purity level.
Note: In each case, purge the line up to the first purifier, bypass the
purifiers, or remove the purifiers during purging. Do not purge the
line through the purifiers or the chromatograph.
Simple Purging Procedure for Fuel Gases
Caution: Be sure to properly vent hydrogen during this stage.
Trained personnel should be present, testing with a portable low
explosive level meter, to ensure that you do not create explosive
concentrations of hydrogen.
Open all valves in the main and branch gas lines. Slowly open the
valve on the cylinder, pressurizing the two-stage regulator. Slowly
open the downstream pressure control knob on the regulator, to
allow gas to flow through the lines at a pressure of 5-10psig. Purge
the lines for 5 minutes, then close the shutoff valves at the vent ends
of the lines. This may involve several points if you have a manifold
system with several branches – startat the branch closest to the
cylinder and work out. Now turn off the cylinder and isolate it by
closing the shutoff valve downstream from the two-stage regulator.
The line is now purged, pressurized, and ready for use. Increase
the pressure in the line to the desired operating pressure (e.g., 40-
60psig).
Purging Carrier and Make-up Gases
Carrier gas and make-up gas systems require a static purge,
followed by a dynamic purge, to ensure the desired purity levels.
Static Purge: Open all valves in the main and branch gas lines, but
close the shutoff valves at the vent ends of the lines. Slowly open
the valve on the cylinder, pressurizing the two-stage regulator.
Slowly open the downstream pressure control knob on the regula-
tor, allowing gas to flow through the lines. Raise the main line
pressure to the pressure you intend to maintain (typically 60-
100psig) Close the downstream pressure control valve on the two-
stage regulator as soon as the pressure is reached. Hold the
system under pressure for 15 minutes, then allow a very small flow
to escape from the shutoff valves at the vent ends of the lines. The
pressure will drop quickly. Close the shutoff valves just before the
pressure reaches zero. It is important that this step not take too
long, or air could leak back into the system. Repeat this step 10
times. This allows impurities trapped in sections of the line which
never get properly swept with gas to diffuse/desorb into the static
purge gas.
Dynamic Purge: After the last static purge, close the shutoff valves
at the vent ends of the lines and bring the main line pressure to
20psig. Choose a shutoff valve as far downstream from the cylinder
as possible. Open this valve slightly, then adjust the valve to allow
a 60cc/min flow of gas. Purge for 24 hours. For this step an extra
flow controller, installed after the opened shutoff valve, will make it
easy to regulate the flow and will help to minimize back diffusion into
the line.
Purifier Connections
After all gas lines are purged it is time to install and purge the
purifiers you will be using. Most Supelco purifiers are factory sealed
with nitrogen or helium and will not contain air. Thus, these devices
require only a short purge time before you can operate your system.
Other purifiers may require many hours to purge. Read the
instructions that accompany the purifiers you intend to use, to be
sure you purge them properly.
At this point all system components should be in place. The only
part of the system left to purge is the short length of tubing
connecting to the chromatograph.
Hereafter, whether the chromatograph is in use or idle, all lines
should remain pressurized at all times.
Installations
Gas chromatograph installations range from simple single-chro-
matograph systems to very complex multi-bench systems. Con-
cerns for a simple installation also are important for the complex
multiple instrument system. If you plan to design a complex system,
you first should read and understand the information presented for
the simpler systems, as well as the basic information in the first
sections of this bulletin. In designing any system, take time to
consider your future needs. Most plumbing problems develop when
a change is made to an existing system. It is good practice to install
your GC with valving and bypass fittings which will quickly allow you
to add one or more GCs. Regardless of how many GCs are involved
the operation, addition, or removal of any GC from the system
should not affect the operation of the other GCs in the system.
SUPELCO
Bulletin 898
14
Installation: Single GC
Location
One of the first steps should be to select a location for the
chromatograph. Site selection is important for many reasons,
including efficient functioning of both the chromatograph and the
operator. Consider temperature and humidity. Generally, instru-
ment manufacturers ask that the room air temperature be between
20°C and 27°C (68-80°F) and the humidity be between 50% and
60% (with no condensation). Air exchange for the oven is very
important to the operation of a GC. The back of the instrument must
be clear for at least 1 foot. The GC will be venting hot air from the
oven to this area. This cannot be accomplished if the vent from one
GC is too close to a wall or the back of another instrument. Don’t
back instruments against one another, or against other heat
sensitive equipment, so that they vent toward each other. Special
vent-directing devices can be installed to avoid these problems
(consult your chromatograph manufacturer). Similarly, don’t place
the GC by a window, or directly under air conditioning or other types
of vents. Don’t place computers, integrators or recorders so that the
paper from these devices is exposed to the vents from ovens or can
spread over heated devices. This can cause a fire or, at the least,
discolor heat-sensitive paper. Ignoring these precautions can
cause erratic temperature control, electrical problems, and shorter
equipment life. For more specific information, consult your instru-
ment manufacturer.
The operator will need room to store samples, along with syringes
and other tools, prior to and after injection. It is best to leave at least
a 2 foot by 2 foot working surface for this purpose. The space
requirement of the GC will be defined by the instrument model, but
in most cases a 3 foot wide space is adequate. Add 2 additional feet
for computer controls and other ancillary devices (autosampler
controls, purge and trap devices, sample concentrators, etc.).
Thus, most GCs and associated devices will require about 6-8
linear feet of counter space. For most labs, this means no more than
3-4 GCs on a 20-24 foot bench.
Gas Cylinders and Gas Lines
Your next decision is to determine where to locate the 1-6 types of
gas cylinders or generators you will need to operate your chromato-
graph. Some facilities prohibit the storage of high pressure cylin-
ders in labs or hallways. Consult your safety department to deter-
mine a suitable location for your cylinders.
Ideally you want the cylinders as close to the GC as possible. The
shortest length of tubing with the fewest connections is best. Never
make a connection in a location that will be hard to access for leak
testing (e.g., overhead, in a ceiling, or behind a bench which is
against a wall).
The diameter of the gas line between the cylinder and the GC
depends on the distance. For a single GC with cylinders within a few
feet of the instrument, 1/16” stainless steel or 1/8” copper tubing
normally are used. These small diameter lines can only be a few
feet long, however, or back pressure will be high (see Table 8). If
the cylinders are further away 1/4“ tubing typically is used. If the
distance is extreme (20 feet or more) 1/2” tubing should be used.
Reduce 1/4" or 1/2" main lines to 1/8" or 1/16" tubing immediately
before the connection to the chromatograph(s).
As a rule, we suggest using larger bore tubing than a first evaluation
would indicate. With larger lines, you have adequate pressure and
flow for additional units, and won’t have to redo the lines. Allow for
expansion and you will save yourself much trouble in the future.
Simple Basic Plumbing For One GC
The diagrams in Figures Q-V show our various alternative recom-
mendations for installing a single GC/FID, using gas cylinders,
cylinders and generators, or generators, and using hydrogen as
fuel or as fuel and carrier gas. TCDs, ECDs, and other detectors do
not require a fuel gas line. Although we recommend gas purifiers as
safeguards, the purifiers shown in Figures Q-V can be removed if
you do not think you need this additional protection. Similarly,
intermediate shutoff valves in the line are useful but not vital. Be
aware that simplifying your system by eliminating purifiers, shutoff
valves, etc. reduces initial costs, but you pay a higher price in terms
of inconvenience (longer downtimes) and loss of column/detector
protection.
The single-cylinder installations in Figures Q-S leave you with the
problem of having to cool down the GC and slowly depressurize the
entire system to change cylinders. Figure V shows a two cylinder
approach which can be used for continuous delivery of any gas.
When the pressure in one cylinder indicates the cylinder must be
changed, the empty cylinder can be closed and the reserve cylinder
opened. We recommend this approach. You should immediately
take the time to change the empty cylinder, however, or the extra
plumbing will be for naught. If you do not change cylinders imme-
diately, chances are good that you will forget to do so, both cylinders
will empty, and you will still have to shut down your GC.
A second approach to cylinder changeover also is viable. An
automatic changeover regulator system (Figure W) connects two
gas cylinders, the active cylinder and a reserve cylinder. When the
pressure in the active cylinder falls below a preset level, gas
automatically begins to flow from the reserve cylinder. You can
change cylinders at your convenience without interrupting the
analysis. The changeover regulator system works on a pressure
differential. The line pressure from the active cylinder is set about
5psig higher than the pressure from the reserve cylinder. Both
cylinders are open, but the reserve cylinder will not deliver gas as
long as the active cylinder can deliver gas at a pressure 5psig higher
than the pressure from the reserve cylinder. This approach requires
two pressure regulators and a downstream in-line regulator, or the
GCs will register the change in pressure when the cylinders switch
operation.
Installation: 2-4 GCs
When you plan to install a 2-4 GC system you must concern
yourself with issues that did not arise with a single GC. Line
diameters and connections, types of purification, valving, locations,
and electrical needs all become more complicated. The plumbing
changes from the relatively simple straight lines of tubing shown in
Figures Q-V, to a complicated assortment of valves, fittings, and
other components. Figure X shows a true manifold system of three
main lines, adequate for 2-4 GCs. For each gas, a two-stage
regulator controls gas pressure in the main line and single-stage
regulators are used in each branch line. We recommend that the
main line pressure be 90-100psig and the individual line regulators
be capable of providing up to 75psig (see Figure F). If your cylinders
are located more than 20 feet from the bench you should use 1/2“
main lines.
Notice that there is a shutoff valve after each branch-off from the
main line. We recommend installing these valves, if your budget
allows, because they enable you to pressure test individual sec-
tions of the system, or isolate each GC and take it off-line without
affecting the operation of the other GCs. The system also
15SUPELCO
Bulletin 898
Figure Q. Ideal Configuration for a Single-GC System: Gas Delivered from Cylinders
Hydrocarbon Trap
Vented to Fume
Hood (or other
approved outlet)
Two-Stage
Pressure
Regulator
Two-Stage
Pressure
Regulator
In-Line Filter
Shutoff
Valve
Shutoff
Valves
Shutoff
Valves
796-0069
Pressure
Gauge
Pressure
Gauge
Shutoff
Valve
Nitrogen
or Helium
Cylinder
(Carrier)
Air
Cylinder
Pressure
Relief
Device
Flashback
Arrestor
Two-Stage
Pressure
Regulator
Pressure
Relief
Device
Pressure
Gauge
Shutoff
Valve
Hydrogen
Cylinder
(Fuel)
GC
In-Line Filter
Hydrocarbon Trap
Hydrocarbon Trap
OMI Tube
In-Line Filter
Molecular Sieve 5A Drying Tube
Pressure
Relief
Device
Shutoff
Valves
Molecular Sieve 5A Drying Tube
has in-line pressure gauges after the two-stage regulators, to
indicate the pressure in the main lines.
In a multiple instrument system, carrier gas and make-up gas
purification also becomes more important. If you have only one
GC, you might not need gas purification, but with all the additional
connections, regulators, and other devices in a system with up to
4 GCs you almost certainly will need several types of purifiers.
Consider the total environment of your multi-GC installation. With
all the integrator and detector cables, gas lines, and electrical
power lines, you will need to allow easy access and identification,
and consider interference with electrical signals. Allow access
room at the front and back of each instrument.
Most commercial laboratory benches are deep enough to install
GCs on both sides, but we recommend splitting the bench,
creating a central access space (not a walkway) to allow access
to the back of each GC. Always stagger the instruments on the
bench, so that heated air from one instrument is not vented
directly at another instrument.
Gas and electric lines should not be left to dangle, this can cause
safety problems and confusion over what gas a line contains.
Various mounting devices are available for gas and power lines,
and these devices should be used. We recommend labels and
color coding for the gas lines. Neatness does count.
SUPELCO
Bulletin 898
16
Figure R. Ideal Configurations for a Single-GC System: Mixed Gas Generator/Gas Cylinder System
Pressure
Relief
Device
Two-Stage
Pressure
Regulator
Hydrogen
Cylinder
(Fuel)
Flashback
Arrestor Hydrocarbon Trap
Shutoff
Valve
Shutoff
Valves
In-Line Filter
Oil Vapor-
Removing
Filter
Oil-Removing
(Coalescing) Filter
Molecular Sieve 5A Drying TubeIn-Line Filter
Zero Air
Generator
Shutoff
Valve
Vented to Fume Hood
(or other approved outlet)
GC
Shutoff
Valve
Pressure
Gauge
In-Line Filter
Shutoff
Valve
Nitrogen
Generator
(Carrier)
In-Line Filter
Pressure
Gauge
OMI Oxygen/Water
IndicatorTube
Shutoff
Valve
Shutoff
Valve
796-0070
Hydrocarbon Trap
Line
(Single-Stage)
Regulator**
Shutoff
Valve
Shutoff
Valve
House
Compressed
Air*
*Replace an oil-sealed compressor with an oilless unit to eliminate the need for the particle filter, oil-removing/coalescing filter, and oil vapor-removing filter.
**Consult generator manual for correct inlet pressure.
Particle
Filter
Line
(Single-Stage)
Regulator**
Pressure
Gauge
Line
(Single-Stage)
Regulator**
17SUPELCO
Bulletin 898
Figure S. Ideal Configurations for a Single-GC System: Hydrogen Used as Carrier and Fuel Gas
GC
Hydrogen
Cylinder
(Fuel &
Carrier)
Two-Stage
Pressure
Regulator PressureRelief
Device
Flashback
Arrestor
Shutoff
Valve
Carrier Gas Line
Shutoff
Valve
Pressure
Gauge Shutoff
Valve
In-Line Filter
OMI Oxygen/Water
IndicatorTubeMolecular Sieve 5A Drying Tube
Shutoff
Valve
Shutoff
Valve
Hydrocarbon
Trap
Pressure
Gauge
796-0071
In-Line Filter
Fuel
Line
Vented to
Fume
Hood (or
other
approved
outlet)
Gas Cylinders or Gas Generators?
With up to 4 GCs you will need many gas cylinders. Carefully select
the site for the cylinders. To keep the instruments running without
interruption, you will not want to shut down instruments to change
cylinders. Thus, additional plumbing (i.e., a two-cylinder system or
an automatic changeover system, as shown in Figures V and W)
will be needed. Furthermore, cylinder changes more frequent than
once per week are an inefficient use of manpower. Calculate your
gas consumption from the equation and example given in the
installation information for 5-20 GC systems. If you will be chang-
ing cylinders more frequently than once per week, consider using
larger cylinders, cradles of cylinders, or gas generators. Cylinder
cradles can be plumbed by your supplier for a single connection to
your system. The location of labs in the centers of buildings often
forces the cylinders-or-generators issue – generators eliminate
the need for very long gas lines or cylinders mounted in hallways.
If you decide to use generators, allow bench or wall space for them,
as near the GCs as possible.
Electrical Concerns
Electrical requirements for installing 2-4 GCs are similar to those
for one GC. Each instrument should be on its own 15-20 amp
circuit. Try to keep related electrical devices (integrator, computer,
etc.), except electrically actuated devices, on the same circuit.
Detector and integrator cables should be shielded and located 6"
or more away from the electrical lines. The gas lines, particularly
copper lines, should be 6-12" away from the power lines – they can
pick up electrical current if they are too close to the power lines.
Of real concern are power interruptions that allow heated zones in
the GCs to cool down. When the power is restored, these heated
zones all come on together and have a tremendous power draw.
When electricity goes off it is best to turn off the main power switch
for each GC. When the electricity comes back on, turn each GC
back on by zones: detector heaters, then the inlet, then the oven.
Consult an electrician about your power needs. Don’t forget to
establish separate earth grounding for your lines.
Hydrocarbon Trap In-Line Filter
Pressure
GaugeAirCylinder
Shut-
off
Valves
Pressure
Relief
Device
Two-Stage
Pressure
Regulator
Shut-
off
Valves
Molecular Sieve 5A Drying Tube
SUPELCO
Bulletin 898
18
Figure T. Ideal Configurations for a Single-GC System: All Generator System
Pressure
Gauge
In-Line Filter
In-Line Filter
Flashback
Arrestor
Molecular Sieve 5A Drying Tube
Vented to Fume Hood
(or other approved outlet)
Shutoff
Valve
Line
(Single-Stage)
Regulator***
In-Line Filter
Shutoff
Valve
Oil-Removing
(Coalescing)
Filter
Oil Vapor-
Removing
Filter
Shutoff
Valve
Shutoff
Valve
Shutoff
Valve
Shutoff
Valves
Pressure
Gauge
Hydrocarbon Trap
OMI Oxygen/Water
IndicatorTube
Shutoff
Valve
Shutoff
Valve
OMI Oxygen/Water
IndicatorTube In-Line Filter
Pressure
Gauge
Shutoff
Valve
Zero Air
Generator
GC
Nitrogen
Generator
(Carrier)
House
Compressed
Air*
Pressure
Relief
Device**
Hydrogen
Generator
(Fuel)
796-0072
**Not needed if the hydrogen generator has a built-in relief device.
***Consult generator manual for correct inlet pressure.
*Replace an oil-sealed compressor with an oilless unit to eliminate the need for the particle filter, oil-removing/coalescing filter, and oil vapor-removing filter.
Particle
Filter
Line
(Single-
Stage)
Regulator***
Line
(Single-Stage)
Regulator***
19SUPELCO
Bulletin 898
Figure U. Ideal Configurations for a Single-GC System:
All Generator System with Hydrogen as Carrier and Fuel Gas
Pressure
Gauge
Flashback
Arrestor
Shutoff
Valve
Pressure
Relief
Device**
GC
Shutoff
Valve
Oil Vapor-
Removing
Filter
Figure W. Automatic Switchover Manifold SystemFigure V. System Using Two Cylinders for Each Gas
Oil-
Removing
(Coalescing)
Filter Hydrocarbon Trap
In-Line Filter
Pressure
Gauge
Shutoff
Valve
Zero Air
Generator
To
Pressure
Relief
Device,
Traps,
etc.
To Pressure
Relief Device,
Traps, etc.
796-0079
796-0257
House
Compressed
Air*
Line
(Single-Stage)
Regulator***
Molecular Sieve 5A Drying TubeIn-Line Filter
Shutoff
Valve
Hydrogen
Generator
Shutoff
Valves
Vented to Fume Hood (or
other approved outlet)
OMI Oxygen/Water
IndicatorTube In-Line Filter
Pressure
Gauge
Shutoff
Valve
Shutoff
Valve
796-0073
<80psi
From
Cylinder
#2
From
Cylinder
#1
85psi 80psi
*Replace an oil-sealed compressor with an oilless unit to eliminate the need for the particle filter, oil-removing/coalescing filter, and oil vapor-removing filter.
**Not needed if the hydrogen generator has a built-in relief device.
***Consult generator manual for correct inlet pressure.
Particle
Filter
Line
(Single-Stage)
Regulator***
SUPELCO
Bulletin 898
20
Figure X. Ideal Configuration for 2-4 GC System (Plumb gas generators as shown in Figure Q)
796-0270
Two-Stage
Pressure
Regulator
Carrier Gas
(Helium or Nitrogen)
Shutoff
Valves
Pressure
Relief
Device
High Capacity
Purifier
Two-Stage
Pressure
Regulator
Vented to
Fume Hood
(or other
approved
outlet)
Shutoff
Valve
Molecular Sieve 5A Drying Tube
Shutoff
Valves
Hydrocarbon Trap
Hydrocarbon Trap
Hydrocarbon Trap
Pressure
Gauge
In-Line Filter
Pressure
Relief
Device
Shutoff
Valves
Two-Stage
Pressure
Regulator
Hydrogen (Fuel Gas)
Pressure
Gauge
In-Line Filter
Pressure
Gauge
In-Line Filter
Pressure
Relief
Device
Flashback
Arrestor
Air
Pressure
Gauge
Molecular Sieve 5A Drying Tube
21SUPELCO
Bulletin 898
796-0271
Shutoff
Valve
Line (Single-Stage)
Pressure Regulator
Shutoff
ValveOMI Tube
Shutoff
Valve
GC
Capped ends, ready for
future expansion
Figure X. (contd.)
SUPELCO
Bulletin 898
22
Figure Y. Complex System in a QA Lab
796-0272
Two-Stage
Pressure
Regulator
Shutoff
Valves
Pressure
Relief
Device
Pressure
Gauge
Shutoff
Valve
Mass Flow Meter
with Alarm Shutoff
Valve
Vented to Fume
Hood (or other
approved outlet)
Shutoff
Valves
Shutoff
Valve
Two-Stage
Pressure
Regulator Flashback
Arrestor
Oil Vapor-
Removing
Filter
Pressure
Relief
Device
Oil-Removing
(Coalescing) Filter
House
Compressed
Air*
Line (SingleStage)
Pressure Regulator
Pressure
Relief
Device
Mass Flow Meter
with Alarm
Hydrogen (Fuel Gas)
Cylinder Cradle
Carrier Gas
Cylinder Cradle
In sequence: shutoff valve, pressure gauge, hydrocarbon trap,
molecular sieve 5A trap, bulk oxygen-water trap, pressure gauge,
in-line filter, shutoff valve.
same sequence as above
same sequence as below
same sequence as above
same sequence as below
same sequence as above
Shutoff
Valve
Shutoff
Valve
Mass Flow Meter
with Alarm
same sequence as below
In sequence: shutoff valve, hydrocarbon trap, molecular sieve 5A
trap, in-line filter, shutoff valve.
In sequence: shutoff valve, pressure gauge,
hydrocarbon trap, molecular sieve 5A trap,
pressure gauge, in-line filter, shutoff valve.
*Replace an oil-sealed compressor with an oilless unit to eliminate the need for the particle filter, oil-removing/coalescing filter, and oil vapor-removing filter.
23SUPELCO
Bulletin 898
Figure Y. (contd.)
796-0273
Line (Single-Stage)
Pressure Regulator
Rotameter
(or Digital Mass Flow Meter)
In-Line Filter
In-Line Filter
Rotameter/Flowmeter
OMI Tube
Shutoff
Valve
1/4" Lines to
Benches 2,3 etc.
1/8" Lines to GCs 2,3 etc. on Bench 1
1/8" Lines to GCs 2,3 etc. on Bench 1
1/8" Lines to GCs 2,3 etc. on Bench 1
1/4" Lines to
Benches 2,3 etc.
1/4" Lines to
Benches 2,3 etc.
1/2" Line
1/2" Line
Bench #1,
Instrument #1
In-Line Filter
1/2" Line
Line (Single-Stage)
Pressure Regulator
Line (Single-Stage)
Pressure Regulator
Rotameter/Flowmeter
SUPELCO
Bulletin 898
24
______________________________ = SCF/day
28,317cc/cubic foot
___________________________ = days/cylinder
SCF/day
Example: nitrogen consumption for 5 GCs
using 266cc nitrogen/min/GC
___________________________ = 67.6 SCF/day
28,317cc/cubic foot
___________________________ = 3.2 days/cylinder
67.6 SCF/day
*Ask your supplier for specifications for the cylinders you use.
Rotameters
In many large-scale GC installations rotameters are used as
visual indicators of gas usage. If the rotameter is of the proper
size, so that gas use per bench suspends the float or ball mid-way
in the rotameter tube, a quick glance will tell you if you are using
the correct amount of gas. Leaks tend to push the float off scale
– they can easily double or more your gas consumption. We
recommend one rotameter for the entire lab and one for each bench
(Figure Y).
Mass flow meters also are used in large facilities, to determine the
total flow of gas into the facility. Often these devices provide an
alarm if the flow is too high. A high flow might indicate a break
somewhere in the lines. The use of rotameters and mass flow
meters to monitor gas consumption can give very good informa-
tion about the integrity of the gas systems, and can help you to
quickly find leaks.
Gas Purifiers
With high gas usage you can consume these devices rapidly. It
might be necessary to mount several purifiers in parallel, to obtain
reasonable life from the individual purifiers (Figure Y). Establish
a maintenance program for regularly changing these purifiers.
x
number
of GCs
mean
flow
(cc/min)
x min/day(60 x 24 = 1440)
SCF/cylinder*
5 GCs x 266cc/min x 1440 min
218 SCF/cylinder*
Installation: 5-20 GCs
The additional major concerns for installing a laboratory of GCs
deal with gas line diameters, gas flow measurement, and upsizing
devices. First, you need to know how much gas the facility might
use. Consider that each GC with two flame-type detectors could
use the amounts of gases listed in Table 2. Add the gas flows for
all the GCs to obtain an estimate of the total gas use. After
converting the volume from cc/min to standard cubic feet (SCF)
per day, divide the volume of gas in one cylinder by the consump-
tion per day. From this calculation you can determine how long
each cylinder should last. Determine this consumption for each of
the gases you intend to use.
A more practical approach is to use larger purifiers. Hydrocarbon
traps and molecular sieve-containing moisture traps with 750cc
adsorbent beds – three to five or more times the capacity of
conventional traps – are available from Supelco. The ¼” or ½”
endfittings on these large traps are compatible with the larger
diameter gas lines used in 5-20 GC systems, minimizing the
pressure drop across the traps. Our large-capacity traps effec-
tively remove contaminants at flow rates of up to 10 liters/minute.
Electrical Considerations
Consult an electrical engineer about the special needs of a large
GC facility. Explain the need for separate, dedicated, grounded
lines for each GC and associated equipment. As when designing
simpler systems, estimate your total power needs by adding
approximately 2100 watts for each GC, and the needs of the
integrators and all peripheral equipment which you anticipate
using.
Additional Resources
From the foregoing information, you can appreciate that there is
latitude in many aspects of installing a gas chromatograph.
Detailed information about in-line gas purifiers is presented in our
Bulletin 848, which is available on request. Specifications and
information about hydrogen generators, nitrogen generators, and
zero air generators are presented in publications 694001, 413065,
and 494053, respectively. The Swagelok Manual (Catalog No.
22339) contains much detailed information about plumbing gas
systems.
As you can appreciate, the information in this bulletin is general in
nature. What is best in most situations may not apply to your
specific situation. If you are unsure about something, it is always
best to ask before proceeding. Supelco technical service and
research chemists are always available to discuss your concerns.
Trademarks
Crescent – Cooper Industries
Drierite – Hammond, W.A., Drierite Company
GateKeeper – Aeronex, Inc.
Glasrench, OMI, Supelcarb, Supelpure – Sigma-Aldrich Co.
GOW-MAC – GOW-MAC Instrument Co.
Hall – Tracor Instruments, Austin, Inc.
Imp – Gould, Inc., Valve & Fittings Div.
Leak-Tec – American Gas & Chemical Co., Ltd.
MAPP – Dow Chemical Co.
Nanochem – Matheson Gas Products
Nupro, Snoop – Nupro Co.
Oxiclear, Oxisorb – MG Industries
Swagelok – Crawford Fitting Co.
Teflon – E.I. du Pont de Nemours & Co., Inc.
Viton – DuPont Dow Elastomers
Whitey – Whitey Co.
For more information, or current prices, contact your nearest Supelco subsidiary listed below. To obtain further contact information, visit our website (www.sigma-aldrich.com), see the Supelco catalog, or contact
Supelco, Bellefonte, PA 16823-0048 USA.
ARGENTINA · Sigma-Aldrich de Argentina, S.A. · Buenos Aires 1119 AUSTRALIA · Sigma-Aldrich Pty. Ltd. · Castle Hill NSW 2154 AUSTRIA · Sigma-Aldrich Handels GmbH · A-1110 Wien
BELGIUM · Sigma-Aldrich N.V./S.A. · B-2880 Bornem BRAZIL · Sigma-Aldrich Quimica Brasil Ltda. · 01239-010 São Paulo, SP CANADA · Sigma-Aldrich Canada, Ltd. · 2149 Winston Park Dr., Oakville, ON L6H 6J8
CZECH REPUBLIC · Sigma-Aldrich s.r.o.· 186 00 Praha 8 DENMARK · Sigma-Aldrich Denmark A/S · DK-2665 Vallensbaek Strand FINLAND · Sigma-Aldrich Finland/YA-Kemia Oy · FIN-00700 Helsinki
FRANCE · Sigma-Aldrich Chimie · 38297 Saint-Quentin-Fallavier Cedex GERMANY · Sigma-Aldrich Chemie GmbH · D-82041 Deisenhofen GREECE · Sigma-Aldrich (o.m.) Ltd. · Ilioupoli 16346, Athens
HUNGARY · Sigma-Aldrich Kft. · H-1067 Budapest INDIA · Sigma-Aldrich Co. · Bangalore 560 048 IRELAND · Sigma-Aldrich Ireland Ltd. · Dublin 24 ISRAEL · Sigma Israel Chemicals Ltd. · Rehovot 76100
ITALY · Sigma-Aldrich s.r.l. · 20151 Milano JAPAN · Sigma-Aldrich Japan K.K. · Chuo-ku, Tokyo 103 KOREA · Sigma-Aldrich Korea · Seoul MALAYSIA · Sigma-Aldrich (M) Sdn. Bhd. · 58200 Kuala Lumpur
MEXICO · Sigma-Aldrich Química S.A. de C.V. · 50200 Toluca NETHERLANDS · Sigma-Aldrich Chemie BV · 3330 AA Zwijndrecht NORWAY · Sigma-Aldrich Norway · Torshov · N-0401 Oslo
POLAND · Sigma-Aldrich Sp. z o.o. · 61-663 Poznañ PORTUGAL· Sigma-Aldrich Quimica, S.A. · Sintra 2710 RUSSIA · Sigma-Aldrich Russia · Moscow 103062 SINGAPORE · Sigma-Aldrich Pte. Ltd.
SOUTH AFRICA · Sigma-Aldrich (pty) Ltd. · Jet Park 1459 SPAIN · Sigma-Aldrich Quimica, S.A. · 28100 Alcobendas, Madrid SWEDEN · Sigma-Aldrich Sweden AB · 135 70 Stockholm
SWITZERLAND · Supelco · CH-9471 Buchs UNITED KINGDOM · Sigma-Aldrich Company Ltd. · Poole, Dorset BH12 4QH
UNITED STATES · Supelco · Supelco Park · Bellefonte, PA 16823-0048 · Phone 800-247-6628 or 814-359-3441 · Fax 800-447-3044 or 814-359-3044 · email:supelco@sial.com H
Supelco is a member of the Sigma-Aldrich family. Supelco products are sold through Sigma-Aldrich, Inc. Sigma-Aldrich warrants that its products conform to the information contained in this and other
Sigma-Aldrich publications. Purchaser must determine the suitability of the product for a particular use. Additional terms and conditions may apply. Please see the reverse side of the invoice or packing slip.
Most popular related searches
