- Temperature :
- Temperature controlled enclosure
- Control range: 20°C to 70°C
- Regulation accuracy: ±0.2°C
- Enclosure also provides anti-condensation protection
- Weight measurement :
- High Resolution microbalance
- Sample mass: between 1 and 1000mg
- Mass change: ±150mg
- Resolution (precision): 0.1 μg
- Peak to peak noise: ≤ 0.3 μg
- Pressure measurement :
- 1000 Torr Baratron full scale from 0.01 up to 1000 Torr (standard)
- 10 Torr Baratron: full scale from 0.005 up to 10 Torr (standard)
- Other ranges available upon request (100 and 500 Torr)
- Accuracy: ±0.5 of Reading
- Resolution: 0.01% of full scale
- Vacuum pressure measurement : Vacuum pressure transducer: full scale from 1×10-8 Torr up to 760 Torr
- Computer hardware and software : The system is fully automated and controlled by purpose-built software. Data can be analyzed in real time using analysis software.
Key measurement capabilities:
- In-situ sample outgassing at temperatures of up to 400 °C and high vacuum (≤10-6 Torr)
- In-situ drying curves of pharmaceutical materials at various pressures and temperatures
- Real-time adsorption / desorption kinetics
- Adsorption / desorption IsoTherms (T constant, P variable)
- Adsorption / desorption IsoBars (T variable, P constant)
- ‘apparent’ BET (Brunauer, Emmett and Teller) Surface Area
- Diffusion coefficients
- Heat of Adsorption
- Dynamic Competitive Adsorption of two probe molecules (e.g. CO2 – H2O, Toluene – H2O, MeOH – H2O)
- Binary competitive adsorption of one probe molecule at constant loading of other probe molecule
- Multiple adsorption/desorption and sample regeneration cycles
DVS Vacuum offers dynamic or static sorption or co-adsorption experiments from low pressures up to 1bar. The system’s key features are precise control and measure of temperature and pressure while recording the changes in sample mass using high-resolution UltraBalace.
DVS Vacuum uses vacuum produced by a combination of roughing and turbomolecular pumps for its operation – in-situ outgassing of sample up to 400 °C and water or organic vapors generation at the set temperature. The vacuum chamber including sample, pressure transducers, vapor/gas delivery system, and solvent flasks are at thermal equilibrium in a single temperature enclosure allowing smooth operation over the entire temperature range (20-70 °C) without risk of condensation. The system introduces partial pressures of water or organic solvent under vacuum to minimize their impact on the structure of a material. The analyzer measures true water or organic vapor sorption isotherms without unnecessary interference from a carrier gas. In addition, it allows us to study the drying kinetics of pharmaceutical powders.
Long Term Temperature Stability
- Vapor generation and delivery manifold at sample temperature
- Prevents condensation issues found in instruments with multiple temperature zones
Pressure measurements
- Direct, Gas Independent, High Accuracy Pressure Measurement
Mass Measurements
- Excellent balance Baseline Stability over 24 hours
- Mass changes at a resolution of 0.01 μg
- Root mean square noise of ≤ 0.3 μg