Frontier Laboratories Ltd

FrontierModel PY-3030S -Single-Shot Pyrolysis System

SHARE

An ideal model for basic polymer research as well as quality control of polymeric products. Single-Shot Pyrolyzer PY-3030S is a pyrolyzer specially designed for flash pyrolysis.

Most popular related searches

High performance and high reliability
Guaranteed reproducibility of polystyrene pyrograms: The reproducibility of the peak area ratio of styrene trimer and methyl stearate internal standard is 2% or less in RSD.

Any type of sample can be analyzed
Samples with a wide range of sample forms can be analyzed using five samplers including two new optional samplers (on-line Micro Reaction Sampler and Micro TD Sampler).

Automated analysis and enhanced reliability
Auto-Shot Sampler (AS-1020E) can be used for automated analysis of up to 48 samples. This improves the reliability of analysis results. entire flow path is deactivated

Furnace temperature control range/stability:
40 - 800 ºC (1 ºC step) / ± 0.1 ºC

Py-GC interface temperature control range/stability:

100 - 400 ºC (1 ºC step) / ± 0.1 ºC

Performance guarantee (polystyrene pyrogram):

Reproducibility of trimmer/internal standard ratio is 2 % or less (RSD)

Requirements:

Gas chromatograph (with split/splitless injector), control PC

Size (furnace):

76 (W) x 143 (D) x 215 (H) mm, 1.6 kg (cables excluded)

Power requirement:

100/120 VAC or 200/240 VAC (400 W max)