ComplianceModel HT-2000S -Hipot Line Tester

SHARE

Hipot Tester, 0-2000Vac @ 20mA and Adjustable Ground Continuity, UL Listed. The newest product of our Hipot Line Testers, the UL Listed HT-S series hipot now comes with an adjustable ground continuity test in range of 0.08 - 1.50 Ω, and Auto-offset for test leads up to 1.00 Ω. Performs 0-2000 AC Hipot Test. Neutral/Hot to Ground Leakage current test up to 20 mA AC. Easy to use and connect. Ships complete with cables, manual, and a Calibration Certificate. Made in the USA. Calibration and repair services available.

Most popular related searches
  • High precision leakage current, duration and ground continuity settings.
  • Allows no load setup of leakage current and test dwell times.
  • Once entered, ground continuity setting, leakage current and test dwell times can be referenced on the front panel display when unit is not testing.
  • Leakage currents can be set from 1-20 mA AC in 0.1mA increments.
  • Dwell times can be set from 1-60 seconds in 1 second increments.
  • Two tests with one setup. Conducts UL/CSA Ground continuity and Hipot tests with one setup, so no additional equipment is needed.
  • Small and light. 9.25" wide x 4.75" high x 10" deep. 11 lbs. But rugged enough to test manufactured housing.
  • Ground continuity check is defeatable for testing two-wire products.
  • Easy equipment connections. Plug your equipment right into the NEMA plug on the front panel. No adapters needed!
  • J2 and J3 cords allow connection of pigtail power leads.
  • Simple test outcome judgement. Green lights mean pass - red lights mean fail. Failing tests sound a buzzer as well.
  • Ready for your UL Inspector out of the box. NIST Traceable Calibration Certificate to MIL-STD45662-A is included. Cal cert is in accordance with UL guidance.
  • Operation Instructions are printed on the rear panel for easy reference.
  • In-depth testing. Both true Arc Dectection and Leakage Current circuits test the product to find problems.
  • RED lights and a buzzer leave no question if a failure has occurred.
  • Expandable for TUV/European Agency Testing. Can be connected to our GF-30ac 25A Ground Bond Tester with optional TEST-LINK™.
  • Easy shift-by-shift Function Test. Compatible with our Patented function checker, the HTT-1and HTT-1R (With 0.1 ohm Ground check Pass/Fail Setup).

  • Voltage Range: 0-2000 AC
  • Leakage Current Set Point: 1-20 mA AC
  • Input Voltage Requirements: 117-127 V, 50 /60 Hz
  • Fuse Requirements: 120 V, 2 A Slow Blow
  • Arc Detect: Independent High Frequency Circuit
  • Leakage Detect: Independent Low Frequency Circuit
  • Ground Continuity Set Point: 0.08 Ω - 1.50 Ω, Defeatable
  • Ground Continuity Test Current: 100 mA
  • Test Time Range: 1 - 60 Seconds
  • Ramp Time Range: 1 - 5 Seconds
  • Voltage Meter Accuracy: ± 2 %
  • HV Shutdown by Ground Fail: < 7 mS
  • Duty Cycle: 100%
  • Discharge Time: < 7 mS VAC
  • Transformer Power: < 500 VA
  • Pass/Fail Repeatability: ± 3 %
  • Output Connector: NEMA plug
  • Display: 4 digits LED display

Test Indicator Led indicators

  • Failure Indicator: Led indicators and buzzer
  • Programable: No
  • PLC Interface: Optional
  • TestLink (With GF-30AC): Optional
  • Computer Control: No
  • Voltage Hold on failure: No

Top Mounted Handle Optional

  • Test Defeat: Ground Continuity
  • Autoreset on Pass: Optional
  • Operation Temperature: 15 - 40 °C
  • Relative Humidity Range: 0 - 90 % Non-condensing
  • Dimensions: 9.25" wide x 4.75" high x 10" deep
  • Weight: 11 Lbs
  • Calibration: One year calibration Certificate
  • Test Cords: Return Lead, Power Cord
  • Test Cords: Ground Continuity Connector Tip
  • Test Cords: J2 and J3 Optional
  • Test Cords: Alligator Clip