PET - Model EQE1800 - External Quantum Efficiency Measurement System
From Spectral Response Systems
Quantum Efficiency (QE) measurements provide a primary characterizationmethod for completed photovoltaic devices. QE system provides electronicsand software designed for fully automated measurement of external quantumefficiency of solar cells. All systems include probes and a fixed plate samplestage for samples up to 150 mm x 150 mm. The main system componentsinclude: custom designed software, measurement electronics, and computersystem (Windows 8 operating system).
Main System Capabilities
The measurement involves focusing monochromatic light to a spot on the device under test, then accurately measuring the photon flux and current from the test cell. The system utilizes a dual beam configuration with lock-in detection, providing an absolute accuracy of ±3%. The QE system uses a grating monochrometer with silicon/lnGaAs/Ge detectors. The system includes automatic order sorting filters and two light sources for monochromatic illumination (a Xenon- arc lamp and a halogen lamp). A single lock-in amplifier is used to measure both the reference detector and test device. The main system comes with all the hardware needed to measure quantum efficiency, a fixed plate sample stage and probes.
Custom designed software using LabView2009 provide...
Custom designed software using LabView2009 provides a user-friendly operation for data collection, plotting and analysis. The software provides flexibility and ease of use in the measurement and manipulation of QE. Full manual control over the mono-chrometer and order sorting filters is provided for diagnostic purposes. The wavelength range and step size during measurements are user definable. Data stored to disc can be directly examined within the supplied software, or can be exported to a text file readable by most third party software packages. Up to six data sets can be plotted on a single graph for comparison. The calculated QE is convoluted with a standard AM1.5 global spectrum to determine the short circuit current.
Options
- Internal Quantum Effeciency Option
A total reflectance measurement using an integrating sphere is OPTIONAL for measurement of Internal Quantum Efficiency (IQE) of a solar cell. The data analysis option calculates IQE= EQE/(1-R), where R is reflection. For monocrystalline silicon based solar cells, this is a must option. - Voltage Biasing
A DC power supply for voltage biasing the test device at up to ±5 V can be included. An electric control box on back of the lock-in-amplifier is included to apply a voltage bias. The voltage bias is very useful sometime to separate individual junctions properly from a multi-junction device. It also helps to look at collection losses in a junction. - White Light Bias And Filter Light Bias Capability
A white light exposure simulates actual solar light conditions in a solar cell whereas filtered light bias at different parts of the solar spectrum helps to isolate contributions from individual cells in a multi-junction cell. - Temperature Control Capability
The temperature stage incorporates solid state thermoelectric detectors to provide a controlled measurement of temperature (±1°C) over 10°C to 60°C over a substrate of size up to 150 mm x 150 mm. The system does include a small pump to keep the stage under vacuum so that one does not have moisture build up at low temperatures. - X-Y Stage Capability Option
An OPTIONAL X-Y stage allows moving of a sample in both directions up to 120 mm distances. The main platform is supported by four re-circulating ball carrier bearings mounted to precisely aligned linear guide rails, which together are capable of providing smooth motion for loads up to 20 kg. A backlash-free lead-screw produces smooth translation that is directly driven by a two-phase stepper motor capable of 25,600 micro-steps per revolution, thus yielding a positional resolution of less than 100 nm. Magnetic limit switches allow homing and overdriving protection in both forward and reverse directions. It is driven by stepper motor controllers, which come with its own dedicated software package. The stage provides micro-stepping resolution of 40 nm and speeds of up to 12 mm/s. - Universal Sample Handling Capability
The stage allows the measurement of a solar cell both in superstrate and substrate structures. - Caliberated Detectors
PET will provide two NIST traceable third party calibrated detectors for reference. Detectors will include interface cables to connect detectors to the QE system.
Customer reviews