Model HVX, HVX-21 - High Voltage (HiPot) Cable & Harness Testers
Containing both low voltage and high voltage subsystems, the automation-ready HVX permits expanded testing for insulation resistance, dielectric breakdown, and Zener diode breakdown voltage, and can be upgraded for 4-Wire measurement and/or Advanced Measurements. The Advanced Measurements option provides increased (x5) 2-wire resistance resolution, and the measurement of backplanes, complex networks, capacitors, wire pair capacitance, twisted pair testing, and cable length. After checking for opens, shorts, miswires, and resistance limits, the system applies a user-selectable voltage to each connection group in the cable - from 10 V to the maximum dc/ac voltage. Ramp Up, Ramp Down, Dwell Time (Test Time), Trip Current, and Trip Delay (Soak Time) are adjustable. Leakage Current detected during the high voltage test phase provides a measure of insulation quality. Insulation resistance up to 1 GΩ or 5 GΩ may be recorded.
-
Most popular related searches
Details
Any leakage current exceeding a preset limit reveals the presence of moisture, flux, or other contamination on exposed contacts.
Low voltage resistance thresholds can be set for contact resistance down to 0.1 Ω, and for isolation up to 5 MΩ. Measure embedded resistors from 100 Ω to 1 MΩ with 1% accuracy, and lesser accuracy from 0.1 Ω to 5 MΩ. LV measurement takes less than 0.5 s.
Robust hardware includes: A TEST pushbutton with READY, PASS, and FAIL indicators for one-button operation; Signal remote control and test probe sockets; External Terminals for basic insulation testing on chassis and individual components; Industry-standard 64- pin dual-row latch headers for easy interface to external test fixtures of your own design for custom applications; USB Communication; and a rugged, 1/16”-thick aluminum case with scratch-proof Lexan surface for long life in an industrial environment.
A simple, in-built scripting capability permits full production-testing automation from test through reports and labelling. Program optional external relay boards to switch diverter gates, lock and release latches, and more. Program Ready, Pass & Fail signals to operate tower lights, LEDs, audible tones, label printers etc. Use optional API and LabView interfaces for integration with other equipment.
HVX series testers allow users to meet industry-standard A620 guidelines. Archival-quality reports show user-selected test parameters such as test voltage, leakage current, and insulation resistance for each wire group, and denote PASS or FAIL.
The system employs numerous measures to ensure operator safety. Each login name must have system administrator authorization to permit enabling of the high voltage function. Once logged in, the operator must confirm awareness of the presence of high voltage by pressing a hardware-enable button for that test session. During testing, a bright red light bar turns on when high voltage is present. If desired, remote test and stop switches are available. Finally, test current that exceeds the maximum 1.5 mA causes an automatic halt of that test and cessation of voltage.
Control Module Price Includes: An expandable 128-point tester; screw terminal board set (CB29, Item 759) or your choice of another board set of equivalent value; installation drive with software and documentation (Getting Started Guide, Software Introduction, User's Manual); two-year Product Support Subscription inclusive of free software upgrades, tech support and warranty. READY TO USE
Select Add-On Options
Hardware: 4-Wire Measurement, Advanced Measurement, HV Remote Control, Expansion Modules
Software: AutoBuild™ Guided Assembly, PinMap™ Fixture Editor, Connector Designer™ Connector Editor, Custom Reporting, Win32/.NET API with LabView™ interface
Accessories: Foot Pedal, Relay Control Board (for external digital control), Protective Cover, Pelican Carrying Case
Customer reviews