The CMP10 is the Spectrally Flat Class A pyranometer with the best price-quality-performance ratio on the market. Based on the established CMP 11 technology, CMP10 extends this quality to applications where maintenance is difficult and/or forms a major part of the cost of ownership. The CMP10 has an internal drying cartridge that will last for at least 10 years if the housing is not opened. This minimizes maintenance significantly. The interval for dome cleaning can be extended, and the quality of measurements maximized, by fitting CMP10 with the recently launched CVF4 ventilation unit. The Pyranometer does not require any power as it supplies a low voltage of 0 to 20 mV in relation to the amount of incoming radiation. When a higher voltage level or a 4 to 20mA signal is required, the AMPBOX is the perfect solution.

Most popular related searches

ISO / IEC classification

ISO 9060 spectrally flat Class A, with ISO / IEC 17025 calibration.

Minimized maintenance

No desiccant change for 10 years, best MTBF with 5 years warranty.

CMP series with the world`s largest installed base

Well known for high quality, durability and accuracy. The CMP pyranometers require no power, so are ideal for remote sites.

Technical Attributes

  • Analog Outputs: 0 - 20 mV
  • Cable Length: 33, 82, 164, 330 ft (10, 25, 50, 100 m)
  • Classification: Spectrally Flat Class A (ISO 9060:2018)
  • Digital Outputs: N.A.
  • Directional Response: ˂ 10 W/m² (up to 80 ° with 1000 W/m² beam)
  • Drying Cartridge and Maintenance Interval: Internal, lasts for 10 years, replaced with every recalibration
  • IP Rating: IP68
  • Irradiance Saturation: 4000 W/m² (Max.)
  • Material Enclosures: Aluminum, anodised
  • Non-linearity: ˂ 0.2 % (0 - 100 W/m²)
  • Non-stability: ˂ 0.5 % (change/year)
  • Operating Humidity: 0 - 100 %
  • Operating Temperature Range: -40 - +176 °F (-40 - +80 °C)
  • Response Time: ˂ 1.7 s (63 %), ˂ 5 s (95 %)
  • Sensitivity: 7 - 14 µV/W/m²
  • Spectral Accuracy: 285 - 2800 nm
  • Temperature Correction: ˂ 1 %: +14 - +104 °F (-10 - +40 °C)
  • Weight: 1.3 lb (600 g)
  • Zero offset A: ˂ 7 W/m²